Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
A selective, sensitive and high-throughput liquid chromatography-tandem mass spectrometry (LC-ESI-MS/MS) method has been developed and validated for the quantitation of Guanfacine in rat plasma. Sample clean-up involved liquid-liquid extraction (LLE) and 100 μL of rat plasma was used. YMC BASIC column (50 mm×2.0 mm, 3.5 µm) was used. Mobile phase used was 10 mM ammonium formate (pH 4.0):acetonitrile (70:30, v/v) at a flow rate of 0.3 mL/min. The parent→product ion transitions for the drug (/ 246.0→159.0) and IS (/ 252.0→161.1) were monitored on a triple quadrupole mass spectrometer, operating in the multiple reaction monitoring (MRM) and positive ion mode. The method was validated over the concentration range of 50.00-10,000.00 pg/mL for Guanfacine. The method was successfully applied into a pharmacokinetic study in rat plasma.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5760995 | PMC |
http://dx.doi.org/10.1016/j.jpha.2013.04.006 | DOI Listing |