Purpose: The purpose of this study was to evaluate ethylene-vinyl acetate (EVA) foam as a new lung substitute in radiotherapy and to study its physical and dosimetric characteristics.
Methods: We calculated the ideal vinyl acetate (VA) content of EVA foam sheets to mimic the physical and dosimetric characteristics of the ICRU lung tissue. We also computed the water-to-medium mass collision stopping power ratios, mass attenuation coefficients, CT numbers, effective atomic numbers and electron densities for: ICRU lung tissue, the RANDO commercial phantom, scaled WATER and EVA foam sheets with varying VA contents in a range between the minimum and maximum values supplied by the manufacturer. For all these substitutes, we simulated percent depth-dose curves with EGSnrc Monte Carlo (MC PDDs) in a water-lung substitute-water slab phantom expressed as dose-to-medium and dose-to-water for 3 × 3- and 10 × 10-cm field sizes. PDD for the 10 × 10-cm field size was also calculated with the MultiGrid Superposition algorithm (MGS PDD) for a relative electron density to water ratio of 0.26. The latter was compared with the MC PDDs in dose-to-water for scaled WATER and EVA foam sheets with the VA content that was most similar to the calculated ideal content that is physically achievable in practice.
Results: We calculated an ideal VA content of 55%; however, the maximum physically achievable content with current manufacturing techniques is 40%. The physical characteristics of the EVA foam sheets with a VA content of 40% (EVA40) are very close to those of the ICRU lung reference. The physical densities of the EVA40 foam sheets ranged from 0.030 to 0.965 g/cm , almost covering the entire physical density range of the inflated/deflated lung (0.260-1.050 g/cm ). Its mass attenuation coefficient at the effective energy of a 6-MV photon beam agrees within 0.8% of the ICRU reference value, and its CT number agrees within 6 HU. The effective atomic number for EVA40 varies by less than 0.42 of the ICRU value, and its effective electron density is within 0.9%. PDDs expressed in dose-to-medium and dose-to-water agree with the ICRU curve within 2% in all regions. PDDs calculated with both MC and MGS were within 1.5%.
Conclusions: The EVA40 is an excellent cork-like lung substitute for radiotherapy applications. From a sole material used in footwear, it is possible to obtain a lung substitute that mimics the physical and dosimetric characteristics of ICRU lung tissue even better than the RANDO commercial phantom.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.12781 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!