Aim: During radiotherapy, in patients with implantable cardioverter-defibrillators (ICDs) malfunctions are considered more likely if doses more than 2 Gy reach the ICD site; however, most malfunctions occur with high-energy (>10 MV) radiations, and the risk is less defined using 6-MV linear accelerators. The purpose of the study is to experimentally evaluate the occurrence of malfunctions in ICDs radiated with a 6-MV linear accelerator at increasing photon doses.

Methods: Thirty-two ICDs from all manufacturers (31 explanted and one demo) were evaluated; all devices with a sufficient battery charge underwent multiple radiations with a 6-MV photon beam reaching a cumulative dose at ICD site of 0.5, 1, 2, 3, 5 and 10 Gy and interrogated after every session. All antitachycardia therapies were left enabled; two ICDs were connected to a rhythm simulator (one simulating a complete atrioventricular block without ventricular activity) and visually monitored by external ECG and the ICD programmer during radiation.

Results: Thirteen ICDs were excluded before radiation because of battery depletion; after radiation up to the cumulative dose at the cardiac implantable electronic device site of 10 Gy, in the remaining 19 devices, programmation and battery charge remained unchanged and no switch to safety mode was observed; oversensing, pacing inhibition or inappropriate antitachycardia therapy were neither recorded nor visually observed during radiation.

Conclusion: With a low-energy accelerator, neither malfunctions nor electromagnetic interferences were detected radiating the ICDs at doses usually reaching the ICD pocket during radiotherapy sessions. In this context, magnet application to avoid oversensing and inappropriate therapy seems, therefore, useless.

Download full-text PDF

Source
http://dx.doi.org/10.2459/JCM.0000000000000623DOI Listing

Publication Analysis

Top Keywords

icd site
8
6-mv linear
8
battery charge
8
cumulative dose
8
site 10 gy
8
icds
6
malfunctions
5
radiotherapy risk
4
risk implantable
4
implantable cardioverter-defibrillator
4

Similar Publications

Background: To investigate cancer incidence and the potential influence of immunosuppressive agents in Korean systemic lupus erythematosus (SLE) patients.

Methods: We conducted a retrospective analysis utilizing data from the Korea Healthcare Bigdata Linked Platform, which integrated the National Central Cancer Registry and National Health Insurance Service databases covering the period 2008-2017. Incidence rates (IRs) per 10,000 person-years (PYs) for site-specific cancers of SLE patients were calculated using ICD-O-3 codes.

View Article and Find Full Text PDF

Combination of immunotherapy and photothermal therapy (PTT) provides a promising therapeutic performance for tumors. However, it still faces negative feedback from suppressive factors such as adenosine. Herein, we developed a new nanodrug that can combine adenosine blockade and ferroptosis to promote the photoimmunotherapy of triple negative breast cancer (TNBC).

View Article and Find Full Text PDF

Studies have demonstrated overall prognostic benefits of ICD implantation in patients at increased risk of sudden cardiac death. However, results are inconsistent in certain subgroups. This study aims to evaluate the prognostic implications of comorbidities on ICD outcomes and compare trends in patient selection and outcomes over a decade-long inclusion period.

View Article and Find Full Text PDF

In situ vaccine (ISV) can activate the anti-tumor immune system by inducing immunogenic cell death (ICD) at the tumor site. However, the development of tumor ISV still faces challenges due to insufficient tumor antigens released by tumor cells and the existence of tumor immunosuppressive microenvironment (TIME). Targeting the STING pathway has been reported to enhance the adjuvant effects of in situ tumor vaccines by initiating innate immunity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!