Real-time volumetric relative dosimetry for magnetic resonance-image-guided radiation therapy (MR-IGRT).

Phys Med Biol

Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States of America. The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, TX, United States of America. The University of Texas MD Anderson Cancer Center, 1400 Pressler Street, Houston, TX 77030, United States of America.

Published: February 2018

The integration of magnetic resonance imaging (MRI) with linear accelerators (linac) has enabled the use of 3D MR-visible gel dosimeters for real-time verification of volumetric dose distributions. Several iron-based radiochromic 3D gels were created in-house then imaged and irradiated in a pre-clinical 1.5 T-7 MV MR-Linac. MR images were acquired using a range of balanced-fast field echo (b-FFE) sequences during irradiation to assess the contrast and dose response in irradiated regions and to minimize the presence of MR artifacts. Out of four radiochromic 3D gel formulations, the FOX 3D gel was found to provide superior MR contrast in the irradiated regions. The FOX gels responded linearly with respect to real-time dose and the signal remained stable post-irradiation for at least 20 min. The response of the FOX gel also was found to be unaffected by the radiofrequency and gradient fields created by the b-FFE sequence during irradiation. A reusable version of the FOX gel was used for b-FFE sequence optimization to reduce artifacts by increasing the number of averages at the expense of temporal resolution. Regardless of the real-time MR sequence used, the FOX 3D gels responded linearly to dose with minimal magnetic field effects due to the strong 1.5 T field or gradient fields present during imaging. These gels can easily be made in-house using non-reusable and reusable formulations depending on the needs of the clinic, and the results of this study encourage further applications of 3D gels for MR-IGRT applications.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aaac22DOI Listing

Publication Analysis

Top Keywords

fox gel
12
irradiated regions
8
fox gels
8
gels responded
8
responded linearly
8
gradient fields
8
b-ffe sequence
8
gel
5
gels
5
fox
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!