Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Changes in the macrogeometry of dental implants are known to influence primary stability and the osseointegration process.
Purpose: The purpose of the present in vitro study was to evaluate the mechanical behavior of geometric changes in the apex region of dental implants.
Methods: Thirty-five cylindrical dental implants (Titamax Ti; Neodent) were machined at the apical third to reproduce the experimental groups: without apical cut (Wc), apical bi-split cut, apical tri-split cut, apical quadri-split cut (Qs). One (control group) (Titamax Ti Ex) (n = 7) without any modifications was added. The implants had the same final dimensions (4.1 x 11 mm ). All implants were inserted into artificial bone blocks and were evaluated by insertion torque and resonance frequency by ISQ values (Osstell). Two-tailed analysis of variance (One-way ANOVA) and Tukey's post-test (P < .05).
Results: Control and Qs implants showed a significant increase of the insertion torque (P < .001). For the resonance frequency, Wc and (control) implants had the greatest ISQ values. However, there's no significant difference between (control) and Qs for the ISQ values (P < .001).
Conclusion: Within the limitations of the present study, the proposed geometries at the apical third of dental implants greatly influenced its insertion torque and primary stability in vitro.