Purpose: The aim of the study was to quantify the isolated tear film adhesion error in a Goldmann applanation tonometer (GAT) prism and in a correcting applanation tonometry surface (CATS) prism.

Methods: The separation force of a tonometer prism adhered by a tear film to a simulated cornea was measured to quantify an isolated tear film adhesion force. Acrylic hemispheres (7.8 mm radius) used as corneas were lathed over the apical 3.06 mm diameter to simulate full applanation contact with the prism surface for both GAT and CATS prisms. Tear film separation measurements were completed with both an artificial tear and fluorescein solutions as a fluid bridge. The applanation mire thicknesses were measured and correlated with the tear film separation measurements. Human cadaver eyes were used to validate simulated cornea tear film separation measurement differences between the GAT and CATS prisms.

Results: The CATS prism tear film adhesion error (2.74±0.21 mmHg) was significantly less than the GAT prism (4.57±0.18 mmHg, <0.001). Tear film adhesion error was independent of applanation mire thickness (=0.09, =0.04). Fluorescein produces more tear film error than artificial tears (+0.51±0.04 mmHg; <0.001). Cadaver eye validation indicated the CATS prism's tear film adhesion error (1.40±0.51 mmHg) was significantly less than that of the GAT prism (3.30±0.38 mmHg; =0.002).

Conclusion: Measured GAT tear film adhesion error is more than previously predicted. A CATS prism significantly reduced tear film adhesion error bŷ41%. Fluorescein solution increases the tear film adhesion compared to artificial tears, while mire thickness has a negligible effect.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5757978PMC
http://dx.doi.org/10.2147/OPTH.S152492DOI Listing

Publication Analysis

Top Keywords

tear film
32
film adhesion
12
film separation
12
tear
9
film
8
quantify isolated
8
isolated tear
8
adhesion error
8
gat prism
8
simulated cornea
8

Similar Publications

: Diabetes is a well-recognised factor inducing a plethora of corneal alterations ranging from dry eye to reduced corneal sensibility, epithelial defects, and reduced cicatrisation. This cohort study aimed to assess the efficacy of a novel ophthalmic solution combining cross-linked hyaluronic acid (CHA), chondroitin sulfate (CS), and inositol (INS) in managing diabetes-induced corneal alterations. Specifically, it evaluated the solution's impact on the tear breakup time (TBUT), the ocular surface disease index (OSDI), and corneal sensitivity after three months of treatment.

View Article and Find Full Text PDF

Dry eye disease (DED) is a multifactorial, chronic, and often relapsing condition with a significant impact on patient quality of life (QoL). Symptoms such as ocular discomfort and visual disturbances are diverse and frequently misaligned with objective clinical signs, complicating diagnosis and management. DED not only interferes with daily activities like reading, driving, and computer use but also imposes a substantial economic burden due to direct healthcare costs and reduced work productivity.

View Article and Find Full Text PDF

Age-related changes to the orbicularis oculi muscle include impaired eyelid function, such as lagophthalmos, alterations in tear film dynamics, and aesthetic changes like wrinkles, festoons, and the descent of soft tissue. To date, the structural and functional changes that would comprehensively increase our understanding of orbicularis aging have not been analyzed. This study aims to investigate functional outcomes using surface electromyography and correlate them with ultrastructural changes in orbicularis during aging.

View Article and Find Full Text PDF

Running-In Behavior and Failure Mechanism Between AgCuNi Alloy and Au-Electroplated Layer.

Sensors (Basel)

December 2024

State Key Laboratory of Tribology in Advanced Equipment, Tsinghua University, Beijing 100084, China.

To avoid wear and tear of the slip ring due to electrical corrosion, the slip ring needs to undergo the running-in process under atmospheric conditions without current after assembly. To address the urgent demand for long-service capability space conductive slip rings in the aerospace field, the running-in behavior and failure mechanism between the AgCuNi alloy and Au-electroplated layer are investigated using a ball-on-disc tribometer in this paper. The results show that the transfer film composed of Au plays an important role in modifying the friction during the sliding process.

View Article and Find Full Text PDF

Perfluorohexyloctane ophthalmic solution (Miebo) and water-free cyclosporine ophthalmic solution 0.1% (Vevye) are recently approved treatments for dry eye disease (DED). Perfluorohexyloctane (PFHO) uses a novel approach to treat evaporative DED, whereas water-free cyclosporine (CsA 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!