Development of an automated flow injection analysis system for determination of phosphate in nutrient solutions.

J Sci Food Agric

Mir Araştırma ve Geliştirme AŞ, Materials and Chemical Research Department, Esenler, İstanbul, Turkey.

Published: August 2018

Background: A fully automated flow injection analysis (FIA) system was developed for determination of phosphate ion in nutrient solutions. This newly developed FIA system is a portable, rapid and sensitive measuring instrument that allows on-line analysis and monitoring of phosphate ion concentration in nutrient solutions. The molybdenum blue method, which is widely used in FIA phosphate analysis, was adapted to the developed FIA system. The method is based on the formation of ammonium Mo(VI) ion by reaction of ammonium molybdate with the phosphate ion present in the medium. The Mo(VI) ion then reacts with ascorbic acid and is reduced to the spectrometrically measurable Mo(V) ion. New software specific for flow analysis was developed in the LabVIEW development environment to control all the components of the FIA system. The important factors affecting the analytical signal were identified as reagent flow rate, injection volume and post-injection flow path length, and they were optimized using Box-Behnken experimental design and response surface methodology.

Results: The optimum point for the maximum analytical signal was calculated as 0.50 mL min reagent flow rate, 100 µL sample injection volume and 60 cm post-injection flow path length. The proposed FIA system had a sampling frequency of 100 samples per hour over a linear working range of 3-100 mg L (R  = 0.9995). The relative standard deviation (RSD) was 1.09% and the limit of detection (LOD) was 0.34 mg L .

Conclusion: Various nutrient solutions from a tomato-growing hydroponic greenhouse were analyzed with the developed FIA system and the results were found to be in good agreement with vanadomolybdate chemical method findings. © 2018 Society of Chemical Industry.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jsfa.8911DOI Listing

Publication Analysis

Top Keywords

fia system
24
nutrient solutions
16
phosphate ion
12
developed fia
12
automated flow
8
flow injection
8
injection analysis
8
determination phosphate
8
movi ion
8
analytical signal
8

Similar Publications

Metabolic profiling of human milk (HM) is indispensable for elucidating mother-milk-infant relationships. We evaluated the Biocrates MxP Quant 500 assay for HM-targeted metabolomics (106 small molecules, 524 lipids) and analyzed in a feasibility test HM from apparently healthy Brazilian mothers (A: 2-8, B: 28-50, C: 88-119 days postpartum, n = 25). Of the 630 possible signatures detectable with this assay, 506 were above the limits of detection in an HM-pool (10 µL) used for assay evaluation, 12 of them above the upper limit of quantitation.

View Article and Find Full Text PDF

Integrated SegFlow, µSIA, and UPLC for Online Sialic Acid Quantitation of Glycoproteins Directly from Bioreactors.

Eng Life Sci

January 2025

Analytical Development & Analytical Attribute Science in Biologics Bristol Myers Squibb Devens Massachusetts USA.

This study emphasizes the critical importance of closely monitoring and controlling the sialic acid content in therapeutic glycoproteins, including EPO, interferon-γ, Orencia, Enbrel, and others, as the level of sialylation directly impacts their pharmacokinetics (PK), immunogenicity, potency, and overall clinical performance due to its influence on protein clearance via hepatic asialoglycoprotein receptors (ASGPR). The ASGPR recognizes and binds to glycoproteins exposed to terminal galactose or N-acetylgalactosamine residues, leading to receptor-mediated endocytosis. Recent studies have demonstrated that sialylation of O-linked glycan plays a role in protecting against macrophage galactose lectin (MGL)-mediated clearance.

View Article and Find Full Text PDF

Advances in the design and application of contactless conductivity detectors for separation, flow-through, microfluidic and sensing techniques: A review.

Anal Chim Acta

February 2025

Department of Hygiene, Third Faculty of Medicine, Charles University, Ruská 87, 100 00, Prague 10, Czech Republic. Electronic address:

The review focuses on the design of detection cells, the use of microcontrollers for processing and evaluation of the detection signal, and the development of multi-detection systems for electromigration, liquid chromatography, flow-through and microfluidic techniques. A separate section is the introduction of modern 3D printing techniques and the use of new printing materials for the design of multidetection systems. In addition to traditional utilisation in separation techniques, new versions of contactless conductivity detectors are finding applications in FIA, SIA, portable and paper based analytical systems or as independent sensors.

View Article and Find Full Text PDF

Rapid identification of asparagine (Asn) deamidation and isoaspartate (Asp) in proteins remains a challenging analytical task during the development of biological therapeutics. For this study, 46 therapeutically relevant peptides corresponding to 13 peptide families (13 unmodified peptides and 33 modified peptides) were obtained; modified peptides included Asn deamidation and isoAsp. The peptide families were characterized by three methods: reversed-phase ultrahigh performance liquid chromatography-mass spectrometry (RP-UHPLC-MS); flow injection analysis high-resolution ion mobility-mass spectrometry (FIA-HRIM-MS); and shortened gradient RP-UHPLC-HRIM-MS.

View Article and Find Full Text PDF

Automated Systems with Fluorescence Detection for Metal Determination: A Review.

Molecules

December 2024

Department of Analytical Chemistry, Institute of Chemistry, Faculty of Science, Pavol Jozef Safarik University in Kosice, Moyzesova 11, 040 01 Kosice, Slovakia.

Industrialization has led to environmental pollution with various hazardous chemicals including pollution with metals. In this regard, the development of highly efficient analytical methods for their determination has received considerable attention to ensure public safety. Currently, scientists are paying more and more attention to the automation of analytical methods, since it permits fast, accurate, and sensitive analysis with minimal exposure of analysts to hazardous substances.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!