Purpose: To determine the prevalence of disposable tonometer versus non-disposable tonometer use in the UK and to determine methods of decontamination and frequency of replacement of prisms.A total of 137 ophthalmology departments were interviewed by telephone using a structured questionnaire. The main outcome measured were:types of tonometer prisms used in clinic (disposable, non-disposable and/or other)average disposable prisms used per clinic sessionaverage lifespan of non-disposable prismsprism preference by glaucoma and other teams within department.A cost and benefit analysis was then performed on the data acquired.

Results: One hundred and fifty-five departments were identified for the survey. Of these, 137 (88.3%) responded. Eighty-one departments (59.1%) used Tonosafe prisms alone, whereas 22 departments (16.1%) used Goldmann non-disposable prisms exclusively. Thirty-five departments (64%) on average have a change rate of 26.5% per year (range: 0-100, median: 20) attributed to damage, loss or theft. Sixteen departments (29%) reported that prisms were used until damaged or lost. Four departments (7%) were uncertain of their prism usage and could not provide further information.

Conclusions: Majority of eye departments in the UK opt for disposable prisms. This survey shows the perceived cost-effectiveness of disposable prisms is overestimated when the true cost of disinfection and damage is taken into account. Significant cost savings coupled with the low risk of infectivity (if decontaminated properly) should prompt clinicians and ophthalmic departments worldwide to reconsider the use of non-disposable prisms.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5721635PMC
http://dx.doi.org/10.1136/bmjophth-2016-000019DOI Listing

Publication Analysis

Top Keywords

disposable prisms
12
prisms
9
departments
9
versus non-disposable
8
non-disposable tonometer
8
tonometer prisms
8
prisms clinic
8
non-disposable prisms
8
disposable
6
non-disposable
6

Similar Publications

Introduction: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GAT) and SUOER SW-500 Rebound Tonometer.

Methods: This was a retrospective observational study where 205 eyes of 106 glaucoma patients had their IOPs measured by 2 fellowship trained ophthalmologists. Data were analyzed using the Bland-Altman method of differences.

View Article and Find Full Text PDF

Optical Quality and Visual Performance With Different Toric Contact Lens Designs.

Eye Contact Lens

November 2023

Department of Innovative Visual Science (S.K.), Osaka University Graduate School of Medicine, Osaka, Japan; Department of Ophthalmology (S.K., N.M., K.N.), Osaka University Graduate School of Medicine, Osaka, Japan; and Kozaki Eye Clinic (N.M., M.T., H.M., R.K., J.K.), Osaka, Japan.

Objectives: To evaluate the optical quality of different toric contact lens (CL) designs and compare their on-eye visual correction performance.

Methods: Twenty soft CL wearers aged 20 to 39 years were enrolled. Two daily disposable silicone-hydrogel toric CLs were tested: the "Eyelid Stabilized Design" (ESD-CL) and prism-ballast design (PB-CL); a spherical daily disposable silicone hydrogel CL (spherical CL) was used as a control.

View Article and Find Full Text PDF

Aim: Comparing intraocular pressure (IOP) measurements using Goldmann applanation prism and TonoSafe® in the population without signs of glaucoma.

Material And Methods: Patients with no ocular pathologies, except ametropia (until ± 4 D) or IOP of <30 mm Hg without signs of glaucoma by optic disc structural analysis by fundus biomicroscopy. The IOP was measured sequentially using the traditional cone and the TonoSafe®, according to a randomization list to determine which device would be used first.

View Article and Find Full Text PDF

A silver-manganese nanocomposite was successfully prepared by the urea hydrolysis method and used to detect chloride ions in sweat electrochemically. The synthesis involves the reaction of manganese sulphate, silver nitrate, and urea at 100 °C for 24 h. The crystalline nature of the particle was studied by diffraction analysis and found to be mixed-phase oxides of manganese alongside the oxides of silver.

View Article and Find Full Text PDF

Background: Successful implementation of endoscopic spinal surgery programs hinges on reliable performance and case cost similar to traditional decompression surgeries of the lumbar spine.

Materials And Methods: To improve the statistical power of studying the durability of endoscopes with routine lumbar endoscopy, the authors performed a retrospective survey study among endoscopic spine surgeons by email and chat groups on social media networks WhatsApp and WeChat. Descriptive and correlative statistics were done on the surgeon's responses recorded in multiple-choice questions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!