Fatigue Failure of External Hexagon Connections on Cemented Implant-Supported Crowns.

Implant Dent

Professor, Department of Biomaterials and Biomimetics, New York University/Hansjörg Wyss Department of Plastic Surgery, NYU, Langone Medical Center/Mechanical and Aerospace Engineering, NYU, Tandon School of Engineering, New York, NY.

Published: April 2018

Purpose: To evaluate the probability of survival and failure modes of different external hexagon connection systems restored with anterior cement-retained single-unit crowns. The postulated null hypothesis was that there would be no differences under accelerated life testing.

Materials And Methods: Fifty-four external hexagon dental implants (∼4 mm diameter) were used for single cement-retained crown replacement and divided into 3 groups: (3i) Full OSSEOTITE, Biomet 3i (n = 18); (OL) OEX P4, Osseolife Implants (n = 18); and (IL) Unihex, Intra-Lock International (n = 18). Abutments were torqued to the implants, and maxillary central incisor crowns were cemented and subjected to step-stress-accelerated life testing in water. Use-level probability Weibull curves and probability of survival for a mission of 100,000 cycles at 200 N (95% 2-sided confidence intervals) were calculated. Stereo and scanning electron microscopes were used for failure inspection.

Results: The beta values for 3i, OL, and IL (1.60, 1.69, and 1.23, respectively) indicated that fatigue accelerated the failure of the 3 groups. Reliability for the 3i and OL (41% and 68%, respectively) was not different between each other, but both were significantly lower than IL group (98%). Abutment screw fracture was the failure mode consistently observed in all groups.

Conclusion: Because the reliability was significantly different between the 3 groups, our postulated null hypothesis was rejected.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ID.0000000000000725DOI Listing

Publication Analysis

Top Keywords

external hexagon
12
probability survival
8
postulated null
8
null hypothesis
8
fatigue failure
4
failure external
4
hexagon connections
4
connections cemented
4
cemented implant-supported
4
implant-supported crowns
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!