Objectives: This study sought to evaluate the Medtronic Evolut PRO Transcatheter Aortic Valve System in patients with severe symptomatic aortic stenosis.

Background: A next-generation self-expanding transcatheter aortic valve was designed with an external pericardial wrap with the intent to reduce paravalvular leak while maintaining the benefits of a low-profile, self-expanding, and repositionable supra-annular valve.

Methods: The Medtronic Evolut PRO Clinical Study included 60 patients undergoing transcatheter aortic valve replacement with the Evolut PRO valve at 8 investigational sites in the United States. Clinical outcomes at 30 days were evaluated using Valve Academic Research Consortium-2 criteria. The 2 primary safety endpoints were the incidence of all-cause mortality at 30 days and the incidence of disabling stroke at 30 days. The primary efficacy endpoint was the proportion of patients with no or trace prosthetic valve regurgitation at 30 days. An independent echocardiographic core laboratory (Mayo Clinic, Rochester, Minnesota) was used to adjudicate all echocardiographic assessments.

Results: All 60 patients received the Evolut PRO valve. At 30 days, 1 patient (1.7%) died and 1 patient (1.7%) experienced a nonfatal disabling stroke. Paravalvular regurgitation at 30 days was absent or trace in 72.4% of patients and was mild in the remainder of patients, with no patients having worse than mild paravavlular leak. The mean atrioventricular gradient was 6.4 ± 2.1 mm Hg and effective orifice area was 2.0 ± 0.5 cm at 30 days.

Conclusions: The safety and efficacy results of this study support the use of the Evolut PRO System for the treatment of severe symptomatic aortic stenosis in patients who are at increased surgical risk, resulting in excellent hemodynamics and minimal paravalvular leak (The Medtronic TAVR 2.0 US Clinical Study; NCT02738853).

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2017.10.014DOI Listing

Publication Analysis

Top Keywords

evolut pro
24
transcatheter aortic
16
aortic valve
16
self-expanding transcatheter
8
valve
8
pericardial wrap
8
medtronic evolut
8
patients
8
severe symptomatic
8
symptomatic aortic
8

Similar Publications

A 90-year-old man received a diagnosis of ascending aortic pseudoaneurysms after transcatheter aortic valve implantation (TAVI) using an Evolut PRO valve (Medtronic). Plug closure of the pseudoaneurysms was successfully performed, and the symptoms improved after the procedure. However, on postoperative day 4, the patient experienced sudden massive hemoptysis and died.

View Article and Find Full Text PDF

Transcatheter Aortic Valve-in-Valve Implantation with Newer Generation Evolut Valve by Size of Failed Bioprosthesis.

Anatol J Cardiol

January 2025

Department of Cardiothoracic Surgery Research, Lankenau Institute for Medical Research, Wynnewood, Pennsylvania, USA ; Department of Cardiothoracic Surgery, Lankenau Heart Institute, Main Line Health Wynnewood, Pennsylvania, USA.

Background: To evaluate the clinical outcomes of valve-in-valve transcatheter aortic valve replacement (ViV TAVR) with newer-generation self-expanding Evolut valves according to the size of the failed surgical bioprosthesis.

Methods: This single-center retrospective study evaluated consecutive patients undergoing ViV TAVR with the Evolut Pro/Pro+/Fx between 2018 and 2022. These patients were compared based on the true internal diameter (ID) of the failed bioprosthesis, specifically ≤19 mm (small group) vs.

View Article and Find Full Text PDF

Background: The self-expanding, supra-annular Evolut valve is an established platform for Transcatheter Aortic Valve Implantation (TAVI). Evolut PRO introduced an outer sealing wrap to mitigate paravalvular leakage. We evaluated the 3-year clinical outcomes and valve performance of the Evolut PRO in standard clinical practice for severe aortic stenosis (AS) patients at intermediate or higher risk for surgery.

View Article and Find Full Text PDF

Background: No data compare newer-generation transcatheter heart valves (THVs) in terms of next-day discharge (NDD) following transfemoral (TF) transcatheter aortic valve implantation (TAVI).

Aims: We aimed to evaluate the safety of NDD in unselected patients who received ACURATE (neo/neo2), Evolut (PRO/PRO+/FX) and the SAPIEN (3/Ultra) THVs.

Methods: This multicentre registry included patients who underwent TF-TAVI without a preprocedural permanent pacemaker implantation (PPI) and were discharged the next day without a new PPI.

View Article and Find Full Text PDF

Background: Head-to-head comparisons of second and third generations of transcatheter heart valves (THVs) are mostly limited to 2-arm studies and to mid-term follow-up. The aim of this study was to simultaneously compare clinical outcomes of transcatheter aortic valve replacement (TAVR) with 4 different THVs at 5 years.

Methods: Patients undergoing transfemoral TAVR with 4 second-generation THV platforms and enrolled in the multicentre prospective OBSERVANT II study from December 2016 to September 2018 were compared according to the THV received.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!