Purpose: To simultaneously image brain lesions and veins in multiple sclerosis.

Methods: An interleaved sequence was developed to simultaneously acquire 3D T2*-weighted (or susceptibility-weighted, SW) and fluid-attenuated inversion recovery (FLAIR) images on a 3.0T MRI system. The pulse sequence parameters were calculated to minimize signal perturbation from steady state while maintaining acceptable image contrast and scan time. Fifteen multiple sclerosis patients were enrolled in this prospective study and underwent a standard MS imaging protocol. In addition, SW and FLAIR images were acquired separately and also in an interleaved manner. The SW and FLAIR images were combined into one image to visualize lesions and penetrating veins. The contrast ratios between white matter lesions and penetrating veins were compared between the interleaved sequence and the individual non-interleaved acquisitions.

Results: Interleaved scanning of the FLAIR and the SW pulse sequences was achieved, producing aligned images, and with similar image contrast as in the non-interleaved images. A total of 1076 lesions were identified in all patients on the combined SW-FLAIR image, of which 968 lesions (90%) had visible penetrating veins. Lesion-to-vein contrast ratio was 32.7 ± 17.9 (mean ± standard deviation) for the interleaved sequence compared to 28.1 ± 13.7 using the separate acquisitions (P<0.001).

Conclusion: The feasibility of interleaved acquisition of susceptibility-weighted and FLAIR images was demonstrated. This sequence provides self-registered images and facilitates the visualization of veins in brain lesions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5980669PMC
http://dx.doi.org/10.1002/mrm.27091DOI Listing

Publication Analysis

Top Keywords

interleaved sequence
12
flair images
12
penetrating veins
12
veins multiple
8
multiple sclerosis
8
image contrast
8
lesions penetrating
8
interleaved
6
flair
5
veins
5

Similar Publications

Pseudo-3D HSQC provides an alternative and easy way to record and analyze quantitative HSQC-data. In the original time-zero extrapolated H-C HSQC (HSQC), three separate 2D constant-time (CT) HSQC-experiments (HSQC, i = 1-3) are acquired, where either 1,2 or 3 consecutive CT-HSQC-propagators are repeated in each pulse sequence, and the 2D integral data from the three 2D experiments is analyzed via linear regression. In the presented pseudo-3D HSQC, HSQC is one of the dimensions and all data is contained within one dataset, which is recorded in interleaved manner by acquiring the same t-value for each HSQC-point before t-incrementation.

View Article and Find Full Text PDF

Background: Cerebrospinal fluid (CSF) motion and pulsatility has been proposed to play a crucial role in clearing brain waste. Although its driving forces remain debated, increasing evidence suggests that large amplitude vasomotion drives such CSF fluctuations. Recently, a fast blood-oxygen-level-dependent (BOLD) fMRI sequence was used to measure the coupling between CSF fluctuations and low-frequency hemodynamic oscillations in the human cortex.

View Article and Find Full Text PDF

This study demonstrates high volume rate bistatic 3-D vascular strain imaging, to overcome well-known challenges caused by the anisotropic resolution and contrast inherent to ultrasound imaging. Approach. Using two synchronized 32x32 element matrix arrays (3.

View Article and Find Full Text PDF

Purpose: To develop and validate a 3D turbo spin-echo (TSE)-compatible approach to enhancing black-blood (BB) effects while preserving T weighting and overall SNR.

Methods: Following the excitation RF pulse, a 180° RF pulse sandwiched by a pair of flow-sensitive dephasing (FSD) gradient pulses in the phase- (y) and partition-encoding (z) directions, respectively, is added. The polarity of FSD gradients in z direction is toggled every TR, achieving an interleaved FSD (iFSD) configuration in y-z plane.

View Article and Find Full Text PDF

Purpose: To perform prospective motion correction (PMC) for improved and susceptibility mapping using a purely navigator-based approach.

Methods: Spherical navigators (SNAVs) were combined with an additional FID readout for simultaneous measurement of motion and zeroth-order field shifts. The resulting FIDSNAVs were interleaved for PMC of a multi-echo gradient echo sequence with retrospective correction.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!