Purpose: To compare knee MRI performed with the integrated parallel acquisition technique (PAT) and simultaneous multislice (SMS) turbo spin echo (TSE) T2-weighted (T2w) sequences with conventional TSE sequences in pediatric patients.

Materials And Methods: This was a retrospective IRB-approved study. Seventy-four subjects (26 male, 48 female, mean age 15.3 years, range 8-20) underwent 3-T MRI of the knee with a T2w TSE pulse sequence prototype with four-fold PAT and SMS acceleration as well as the standard PAT-only accelerated sequences. Images were anonymized and two study folders were created: one examination with only T2w PAT2 images (conventional examination) and one examination with only T2w SMS2/PAT2 sequences (SMS examination). Two readers rated examinations for 15 specific imaging findings and 5 quality metrics. Interreader agreement was measured. Signal to noise (SNR) and contrast to noise (CNR) were measured for SMS and conventional T2w sequences.

Results: Consensus review demonstrated diagnostic quality performance of SMS examinations with respect to all 15 structures. Average area under the curve (AROC) was 0.95 and 0.97 for readers 1 and 2, respectively. The conventional sequence was favored over SMS for four out of five quality metrics (p < 0.001). SNR and CNR were higher for the conventional sequences compared to SMS.

Conclusion: SMS accelerated T2w TSE sequences offer a faster alternative for knee imaging in pediatric patients without compromise in diagnostic performance despite diminished SNR. The four-fold acceleration of SMS is beneficial to pediatric patients who often have difficulty staying still for long MRI examinations.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00256-017-2868-2DOI Listing

Publication Analysis

Top Keywords

turbo spin
8
spin echo
8
examination t2w
8
quality metrics
8
sms
6
t2w
5
simultaneous multi-slice
4
multi-slice accelerated
4
accelerated turbo
4
echo knee
4

Similar Publications

Comparison of DWI techniques in patients with epidermoid cyst: TGSE-BLADE DWI vs. SS-EPI DWI.

Jpn J Radiol

December 2024

Department of Diagnostic Imaging and Nuclear Medicine, Graduate School of Medicine, Kyoto University, 54 Shogoin Kawaharacho, Kyoto, 6068507, Japan.

Purpose: To compare quantitative values and image quality between single-shot echo-planar imaging (SS-EPI) diffusion-weighted imaging (DWI) and two-dimensional turbo gradient- and spin-echo DWI with non-Cartesian BLADE trajectory (TGSE-BLADE DWI) in patients with epidermoid cyst.

Methods: Patients with epidermoid cyst who underwent both SS-EPI DWI and TGSE-BLADE DWI were included in this study. Two raters placed ROIs encircling the entire epidermoid cyst on SS-EPI DWI, and then on TGSE-BLADE DWI.

View Article and Find Full Text PDF

Objectives: To explore texture analysis' ability on T and T relaxation maps to classify liver fibrosis into no-to-mild liver fibrosis (nmF) versus severe fibrosis (sF) group using machine learning algorithms and histology as reference standard.

Materials And Methods: In this single-center study, patients undergoing 3 T MRI who also had histology examination were retrospectively enrolled. SNAPSHOT-FLASH sequence for T1 mapping, radial turbo-spin-echo sequence for T2 mapping and spin-echo echo-planar-imaging magnetic resonance elastography (MRE) sequences were analyzed.

View Article and Find Full Text PDF

Background: Traditional echo-planar imaging with intravoxel incoherent motion (EPI-IVIM) exhibits significant magnetic susceptibility artifacts and geometric distortions, which limits its application in nasopharyngeal carcinoma (NPC). This study aimed to compare the qualitative and quantitative indicators between turbo spin echo with intravoxel incoherent motion (TSE-IVIM) and EPI-IVIM in patients with NPC and to provide optimal scanning strategies based on the relationships among these indicators.

Methods: A cross-sectional study was conducted between March 2022 and August 2022.

View Article and Find Full Text PDF

Background: When antispasmodics are unavailable, the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER; called BLADE by Siemens Healthineers) or half Fourier single-shot turbo spin echo (HASTE) is clinically used in gynecologic MRI. However, their imaging qualities are limited compared to Turbo Spin Echo (TSE) with antispasmodics. Even with antispasmodics, TSE can be artifact-affected, necessitating a rapid backup sequence.

View Article and Find Full Text PDF

FeCl and GdCl solutions as superfast relaxation modifiers for agarose gel: a quantitative analysis.

MAGMA

December 2024

Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS Sukolilo, Surabaya, 60111, Indonesia.

Object: This study aimed to evaluate the relaxivity and uniformity of agarose gel phantoms added with relaxation modifiers. It is hypothesized that the modifiers could manipulate the T1 and T2 relaxations as well as the signal uniformity.

Materials And Methods: Twenty agarose gel phantoms with different GdCl₃ and FeCl₃ volume fractions were prepared.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?