Validation of MRI to TRUS registration for high-dose-rate prostate brachytherapy.

Brachytherapy

Département de radio-oncologie, CHUM, Montréal, Québec, Canada; CRCHUM, Université de Montréal, Montréal, Québec, Canada; TECHNA Institute, Toronto, Ontario, Canada. Electronic address:

Published: January 2019

Purpose: The objective of this study was to develop and validate an open-source module for MRI to transrectal ultrasound (TRUS) registration to support tumor-targeted prostate brachytherapy.

Methods And Materials: In this study, 15 patients with prostate cancer lesions visible on multiparametric MRI were selected for the validation. T2-weighted images with 1-mm isotropic voxel size and diffusion weighted images were acquired on a 1.5T Siemens imager. Three-dimensional (3D) TRUS images with 0.5-mm slice thickness were acquired. The investigated registration module was incorporated in the open-source 3D Slicer platform, which can compute rigid and deformable transformations. An extension of 3D Slicer, SlicerRT, allows import of and export to DICOM-RT formats. For validation, similarity indices, prostate volumes, and centroid positions were determined in addition to registration errors for common 3D points identified by an experienced radiation oncologist.

Results: The average time to compute the registration was 35 ± 3 s. For the rigid and deformable registration, respectively, Dice similarity coefficients were 0.87 ± 0.05 and 0.93 ± 0.01 while the 95% Hausdorff distances were 4.2 ± 1.0 and 2.2 ± 0.3 mm. MRI volumes obtained after the rigid and deformable registration were not statistically different (p > 0.05) from reference TRUS volumes. For the rigid and deformable registration, respectively, 3D distance errors between reference and registered centroid positions were 2.1 ± 1.0 and 0.4 ± 0.1 mm while registration errors between common points were 3.5 ± 3.2 and 2.3 ± 1.1 mm. Deformable registration was found significantly better (p < 0.05) than rigid registration for all parameters.

Conclusions: An open-source MRI to TRUS registration platform was validated for integration in the brachytherapy workflow.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.brachy.2017.11.018DOI Listing

Publication Analysis

Top Keywords

rigid deformable
16
deformable registration
16
registration
12
trus registration
12
mri trus
8
centroid positions
8
registration errors
8
errors common
8
common points
8
volumes rigid
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!