Purpose: Use standardized methods to determine how assessment of protective value of radiation-protective garments changes under conditions employing standard beam qualities, scatter-mimicking primary beams, and a modified H (10) measurement.

Methods: The shielding properties of radiation-protective garments depend on the spectrum of beam energies striking the garment and the attenuation properties of materials used to construct the garment, including x-ray fluorescence produced by these materials. In this study the primary beam spectra employed during clinical interventional radiology and cardiology procedures (clinical primary beams, CPB) were identified using radiation dose structured reports (RDSR) and fluoroscope log data. Monte Carlo simulation was used to determine the scattered radiation spectra produced by these CPB during typical clinical application. For these scattered spectra, scatter-mimicking primary beams (SMPB) were determined using numerical optimization-based spectral reconstruction that adjusted kV and filtration to produce the SMPB that optimally matched the scattered spectrum for each CPB. The penetration of a subset of SMPB through four radiation-protective garments of varying compositions and nominal thicknesses was measured using a geometry specified by the International Electrotechnical Commission (IEC). The diagnostic radiological index of protection (DRIP), which increases with increasing penetration through a garment, was calculated using these measurements. Penetration through the same garments was measured for standard beams specified by the American Society of Testing and Materials (ASTM). Finally, 10 mm of PMMA was affixed to the inside of each garment and the DRIP remeasured in this configuration to simulate H (10).

Results: The SMPB based on actual CPB were in general characterized by lower kV (range 60-76) and higher half-value layer (HVL, range 3.44-4.89 mm Al) than standard beam qualities specified by ASTM (kV range 70-85; HVL range 3.4-4.0 mm Al). A lead garment of nominal thickness 0.5 mm (D) had a DRIP of 0.8%, two lead-free garments of 0.5 mm nominal thickness had DRIPs of 1.2% (A) and 2.2% (B), and a lead-free bilayer (C) had a DRIP of 1.4%. When standard beam qualities specified by the ASTM were used, the DRIP for D was 2.2%, 175% higher than the DRIP measured using SMPB, and for A, B, and C was 2.8%, 3.2%, and 2.9%, respectively. This was 133%, 45%, and 107% higher than the DRIP measured using SMPB. Differences between the DRIP of lead-alternative garments and the lead garment were reduced when measured with 10 mm of PMMA. Using this method, the measured DRIPs were 2.2% (A), 3.1% (B), 2.5% (C), and 2.3% (D).

Conclusions: Penetration of radiation through radiation-protective garments depended strongly on the methods and X-ray spectra used for evaluation. The DRIP was higher (i.e., protective value was lower) for lead-alternative garments than for lead garments in this evaluation. The DRIP was lower for all garments when SMPB based on actual clinical beam quality data were used to measure penetration compared to ASTM standard beams. Differences in penetration between lead-alternative and lead garments were less when the DRIP was measured with 10 mm of PMMA between the garment and the chamber.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.12743DOI Listing

Publication Analysis

Top Keywords

radiation-protective garments
20
standard beams
12
garments
12
standard beam
12
beam qualities
12
primary beams
12
10 mm pmma
12
drip measured
12
drip
11
measure penetration
8

Similar Publications

Development and performance evaluation of medical radiation-reducing creams using eco-friendly radiation-shielding composites.

Sci Rep

September 2024

Department of Biomedical Engineering, Keimyung University Department of Medical Informatics, School of Medicine, Keimyung University, Daegu, Korea.

To ensure the safety of medical personnel in healthcare organizations, radiation-shielding materials like protective clothing are used to protect against low-dose radiation, such as scattered rays. The extremities, particularly the hands, are the most exposed to radiation. New materials that can be directly coated onto the skin would be more cost-effective, efficient, and convenient than gloves.

View Article and Find Full Text PDF

[Evaluation of Transmitted X-ray Spectrum, Lead Equivalent, and Uniformity of Radiation Protective Clothing Made of Lead-containing and Lead-free Materials].

Nihon Hoshasen Gijutsu Gakkai Zasshi

September 2024

Division of Medical Quantum Science, Department of Health Science, Faculty of Medical Science, Kyushu University.

Purpose: The purpose of this study was to evaluate the protective performance of several new radiation-protective clothing and to clarify issues of quality control.

Methods: The composition of the shielding elements was analyzed using X-ray fluorescence analysis, and the energy spectrum of transmitted X-rays was measured. Furthermore, the lead equivalent and uniformity were measured from the transmitted X-ray doses according to Japanese industrial standards (JIS).

View Article and Find Full Text PDF

Recently, investigation of advanced shielding materials to be used as an alternative to lead apron has become important. In the current study, MgO loaded into PVC matrix as a non-lead modern shielding composite was modeled to evaluate its performance on radiation protective clothing (RPC). Parameters such as mass attenuation coefficient (MAC), mean free path (MFP), flux buildup factor (FBF), transmission factor (TF) and lead equivalent value (LEV) of samples were calculated using MCNPX Code.

View Article and Find Full Text PDF

Prediction Model for Defects in Lead and Lead-free Aprons.

Health Phys

June 2024

Department of Data Analysis, Faculty of Psychology and Educational Sciences, Ghent University, 9000 Ghent, Belgium.

Personal radiation protective equipment (PRPE) is prone to defects in the attenuating layers, resulting in inadequate protection. Hence, quality control (QC) of PRPE is needed to assess its integrity. Unfortunately, QC of PRPE is laborious and time consuming.

View Article and Find Full Text PDF

Purpose: Several studies report occupational orthopedic problems among interventional cardiologists. These health problems are usually multifactorial. However, the personal protective equipment used should play a major role.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!