Purpose: To describe a modified up-and-over access technique for treatment of iliac artery aneurysms in patients with prior bifurcated stent-grafts for endovascular aneurysm repair (EVAR).

Technique: This technique uses a coaxial 12-F flexible sheath that is docked with a through-and-through wire into a 7-F sheath advanced from the contralateral femoral approach. This maneuver allows both sheaths to be moved as a unit while maintaining position of the apex of the system as it loops over the flow divider, avoiding damage to or displacing the extant endograft. Once the 12-F sheath is positioned in the iliac limb of the aortic stent-graft and secured in place with the through-and-through wire, the repair is extended into the internal iliac artery using a bridging stent-graft or covered stent introduced via a coaxial sheath.

Conclusion: The up-and-over technique with a flexible 12-F sheath mated with a 7-F sheath from the opposite side allows bilateral femoral access to be used for iliac branch device placement after prior aortic endograft procedures that create a higher, acutely angled bifurcation. Use of a through-and-through wire and a coaxial sheath for stent delivery creates a very stable platform for intervention.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1526602817747283DOI Listing

Publication Analysis

Top Keywords

through-and-through wire
12
up-and-over technique
8
iliac branch
8
prior aortic
8
aortic endograft
8
iliac artery
8
7-f sheath
8
12-f sheath
8
sheath
6
iliac
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!