Purpose: The GoCheck Kids smartphone photoscreening app (Gobiquity Mobile Health, Scottsdale, Arizona, USA), introduced in 2014, is marketed to pediatricians with little published validation. We wished to evaluate the GoCheck Kids Screener for accuracy in detecting amblyopia risk factors (ARF) using 2013 American Association for Pediatric Ophthalmology and Strabismus guidelines.

Design: Validity assessment.

Methods: Children 6 months to 6 years of age presenting from October 2016 to August 2017 were included. Children were screened with the GoCheck preloaded Nokia Lumia 1020, software version 4.6 with image processing version R4d, prior to undergoing a comprehensive eye examination by a pediatric ophthalmologist masked to the screener results. Determination of the presence of age-specific ARF was made based upon the examination and compared with the GoCheck recommendation.

Results: A total of 206 children were included (average age 43 months). When compared to examination, GoCheck had a sensitivity of 76.0% and specificity of 67.2% in detecting ARF. Positive predictive value was 57.0% and negative predictive value 83.0%. The screener results of 13 children were changed from "no risk factors" to "risk factors identified" based on the GoCheck remote review process. Four images remained "not gradable" and screening was unsuccessful in 3 children.

Conclusion: In our high-risk population, this version of the Gocheck Kids smartphone app was useful in identifying ARF in children who are often not able to cooperate with visual acuity testing. This study informs pediatricians about the efficacy of this new screener as they make decisions about how to best detect vision problems in young children.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2017.12.020DOI Listing

Publication Analysis

Top Keywords

gocheck kids
16
detecting amblyopia
8
amblyopia risk
8
risk factors
8
kids smartphone
8
gocheck
7
children
6
screener
5
effectiveness gocheck
4
kids
4

Similar Publications

Purpose: To compare the Spot Vision Screener (SPOT) to the GoCheck Kids mobile application (GCK) in the evaluation of amblyopia risk factors (ARFs), according to updated 2021 AAPOS guidelines for instrument-based pediatric vision screening.

Methods: SPOT, GCK using horizontal and vertical images (two-photo mode), and a complete eye examination (CEE) were performed on children ages ≥1 to <7 years referred to a pediatric ophthalmologist from October 2020 to December 2022 for a failed vision screen based on manufacturer guidelines. The primary outcomes were sensitivity, specificity, and positive predictive value (PPV) for each photoscreener.

View Article and Find Full Text PDF

Purpose: To assess the validity of the GoCheck Kids photoscreening application (Gobiquity Mobile Health, Scottdale, AZ) on iPhone, which was used (2018-2022) as standard of care by Child and Family (Kind en Gezin) to detect amblyopia risk factors in children 12-30 months of age.

Methods: Between August 2021 and May 2022, 453 children 11-16 months of age underwent a confirmatory ophthalmic examination within 2 months of GoCheck Kids photoscreening at Child and Family, Flanders, Belgium. Additionally, manual review was performed by specialists of GoCheck Kids.

View Article and Find Full Text PDF

Introduction: Numerous studies have demonstrated the use of artificial intelligence (AI) for early detection of referable diabetic retinopathy (RDR). A direct comparison of these multiple automated diabetic retinopathy (DR) image assessment softwares (ARIAs) is, however, challenging. We retrospectively compared the performance of two modern ARIAs, IDx-DR and Medios AI.

View Article and Find Full Text PDF

Comparative Validation of PlusoptiX and AI-Optic Photoscreeners in Children with High Amblyopia Risk Factor Prevalence.

Clin Ophthalmol

August 2022

Alaska Blind Child Discovery, Alaska Children's EYE & Strabismus, Anchorage, AK, USA.

Purpose: In 2022, an inexpensive multi-radial infrared photoscreener, the AI Optic was released in a similar format as the 2012 PlusoptiX a-12 but utilizing artificial intelligence (AI) for online, central image interpretation. We studied them because no prior comparative validation concerning amblyopia risk factors and particularly refractive error has been done.

Patients And Methods: Children from a pediatric ophthalmology practice had AI Optic and PlusoptiX-a12 photoscreen concomitantly during comprehensive examination with precisely measured strabismus and refraction.

View Article and Find Full Text PDF

The impact of autostereoscopic screens on assessing depth perception.

Ann Transl Med

June 2022

Department of Alaska Blind Child Discovery, Children's EYE & Strabismus, Anchorage, AK, USA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!