A treatment planning comparison between a novel rotating gamma system and robotic linear accelerator based intracranial stereotactic radiosurgery/radiotherapy.

Phys Med Biol

Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia PA, United States of America. The first two authors contributed equally to the study.

Published: February 2018

To compare the dosimetric parameters of a novel rotating gamma ray system (RGS) with well-established CyberKnife system (CK) for treating malignant brain lesions. RGS has a treatment head of 16 cobalt-60 sources focused to the isocenter, which can rotate 360° on the ring gantry and swing 35° in the superior direction. We compared several dosimetric parameters in 10 patients undergoing brain stereotactic radiosurgery including plan normalization, number of beams and nodes for CK and shots for RGS, collimators used, estimated treatment time, D 2 cm and conformity index (CI) among two modalities. The median plan normalization for RGS was 56.7% versus 68.5% (p  =  0.002) for CK plans. The median number of shots from RGS was 7.5 whereas the median number of beams and nodes for CK was 79.5 and 46. The median collimator's diameter used was 3.5 mm for RGS as compared to 5 mm for CK (p  =  0.26). Mean D 2 cm was 5.57 Gy for CyberKnife whereas it was 3.11 Gy for RGS (p  =  0.99). For RGS plans, the median CI was 1.4 compared to 1.3 for the CK treatment plans (p  =  0.98). The average minimum and maximum doses to optic chiasm were 21 and 93 cGy for RGS as compared to 32 and 209 cGy for CK whereas these were 0.5 and 364 cGy by RGS and 18 and 399 cGy by CK to brainstem. The mean V12 Gy for brain predicting for radionecrosis with RGS was 3.75 cm as compared to 4.09 cm with the CK (p  =  0.41). The dosimetric parameters of a novel RGS with a ring type gantry are comparable with CyberKnife, allowing its use for intracranial lesions and is worth exploring in a clinical setting.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aaa517DOI Listing

Publication Analysis

Top Keywords

dosimetric parameters
12
rgs
12
novel rotating
8
rotating gamma
8
parameters novel
8
plan normalization
8
number beams
8
beams nodes
8
shots rgs
8
plans median
8

Similar Publications

Purpose: Increasing life expectancy and advances in cancer treatment will lead to more patients needing both radiation therapy (RT) and cardiac implantable electronic devices (CIEDs). CIEDs, including pacemakers and defibrillators, are essential for managing cardiac arrhythmias and heart failure. Telemetric monitoring of CIEDs checks battery status, lead function, settings, and diagnostic data, thereby identifying software deviations or damage.

View Article and Find Full Text PDF

Purpose: Our objective was to identify the dosimetric parameters and prostate volume that most accurately predict the incidence of acute and late gastrointestinal (GI) and genitourinary (GU) toxicity in prostate cancer stereotactic ablative radiotherapy (SABR) treatments.

Methods: We conducted a retrospective analysis of 122 patients who received SABR for prostate cancer at our clinic between March 2018 and September 2022 using a five-fraction SABR regimen. The existing plans of these patients were re-evaluated according to our institutional protocols (Hacettepe University [HU-1] and HU-2) as well as PACE‑B, RTOG 0938, and NRG GU005 dose-volume constraints.

View Article and Find Full Text PDF

Aim: To characterize the differences of dynamic changes for absolute lymphocyte count (ALC) among esophageal squamous cell carcinoma (ESCC) patients treated with neoadjuvant chemoradiotherapy (nCRT) with or without pembrolizumab, as well as to investigate the clinical and lymphocyte-related organs dosimetric parameters that would impact ALC nadir during nCRT.

Materials And Methods: A total of 216 ESCC patients who received nCRT (with pembrolizumab 144; without pembrolizumab: 72) were identified from a prospective cohort. Weekly and 1-month post-nCRT ALC were identified.

View Article and Find Full Text PDF

Purpose: To investigate the performance of a machine learning-based segmentation method for treatment planning of gastric cancer.

Materials And Methods: Eighteen patients planned to be irradiated for gastric cancer were studied. The target and the surrounding organs-at-risk (OARs) were manually delineated on CT scans.

View Article and Find Full Text PDF

Background: This study aimed to investigate the effect of couch rotation angles on non-coplanar volumetric modulated arc therapy (ncVMAT) plan for stereotactic body radiotherapy (SBRT) in lung cancer patients and to evaluate the feasibility of clinically applying ncVMAT for SBRT.

Methods: Twenty-four lung cancer patients with a single lesion eligible for SBRT were enrolled in the study. Seven dual partial-arc VMAT plans with varying couch angles were designed for every patient.