Purpose: Analysis of intrafraction motion in patients with intracranial targets treated with frameless, mask based stereotactic radiosurgery / radiotherapy using standard couch and 6D-skull tracking on CyberKnife.

Materials And Methods: Twenty-seven treatment datasets of fifteen patients were analyzed. For each sequential pair of images, the correction to the target position (position "offset") in six-degrees of motion was obtained. These offsets were used to calculate intrafraction shifts, and their statistical distribution. PTV margins were calculated, based on Van Herk formula.

Results: The mean ± 1 SD intrafraction translationals were 0.27±0.61mm in left-right, 0.24±0.62mm in antero-posterior and 0.14±0.24mm in supero-inferior direction, and rotations were 0.13±0.21 degrees roll, 0.18±0.25 degrees pitch and 0.28±0.44 degrees yaw. Most intrafraction shifts were ≤ 1mm and 1 degree. Fourteen instances of intrafraction shifts exceeding the robotic correction threshold were noted. Calculated PTV margins were 1mm, 1mm and 0.4mm for for left-right, antero-posterior and supero-inferior directions, respectively.

Conclusions: CyberKnife 6D-skull tracking with 1mm PTV margin effectively compensates for intrafraction motion. The occasional large intrafraction movements may assume significance for techniques not employing intrafraction motion management.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5658803PMC

Publication Analysis

Top Keywords

intrafraction motion
16
6d-skull tracking
12
intrafraction shifts
12
analysis intrafraction
8
mask based
8
intrafraction
8
ptv margins
8
motion
5
motion cyberknife-based
4
cyberknife-based stereotaxy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!