Introduction: Transradial coronary angiography (TRC) can be performed with a one-catheter approach for the right and left coronary ostium (R/LCO). We investigated the performance of a special diagnostic catheter widely used for the one-catheter-approach, the Tiger (Tiger II, TerumoTM).

Methods: In a dual center registry we analyzed 1412 TRC-procedures exclusively performed by experienced TRC-operators. We compared the performance of the Tiger with Judkins catheters by retrospectively judging ostial catheter stability during contrast injection, and by measuring contrast use, fluoroscopy time (FT) and complication rate.

Results: Poor or failed ostial engagement was found in 40.5% in the Tiger group, compared to 46.6% with the use of Judkins catheters (p<0.183). Ostial instability of the Tiger was found more often during engagement of the LCO than the RCO (34.4% vs. 10.8%, p<0.001), whereas it was similar in the LCO and RCO for Judkins catheters (27.4% vs. 26.7%, p = 0.840). TRC-procedures performed with Tiger catheters were associated with less contrast volume (63.48 ± 29.83mL vs. 82.51 ± 56.58mL, p<0.004) and shorter FT than with Judkins catheters. (198.27 ± 194.8sec vs. 326.85 ± 329.70sec). Forearm hematomas occurred less often with the Tiger (1.2% vs. 6.6%, p< 0.02).

Conclusion: The Tiger employed as a single catheter in TRC is an effective tool to achieve lower contrast volume and fluoroscopy time at a low complication rate. Unstable engagement affects predominantly the left coronary artery, but its overall frequency is similar for both, the Tiger and Judkins catheters.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5749733PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0189899PLOS

Publication Analysis

Top Keywords

transradial coronary
8
coronary angiography
8
judkins catheters
8
efficacy one-catheter
4
one-catheter concept
4
concept transradial
4
angiography introduction
4
introduction transradial
4
angiography trc
4
trc performed
4

Similar Publications

Background: Trans-radial coronary angiogram (TR-CAG) has gained popularity due to lower complication rates compared to transfemoral access. Operators can use either conventional catheters, such as Judkins, or single dedicated catheters, such as Tiger-II. This meta-analysis compared the safety and efficacy of Tiger-II versus Judkins catheters in TR-CAG.

View Article and Find Full Text PDF

Aims: This study aims to verify the feasibility and safety of percutaneous coronary intervention (PCI) after a distal transradial approach (dTRA) with radial artery occlusion (RAO) recanalization.

Methods: Between July 2018 and January 2022, 30 patients underwent PCI following attempted RAO recanalization via dTRA. Among these cases, the target radial arteries could not be recanalized in five patients, necessitating alternative vascular access.

View Article and Find Full Text PDF

Vascular access: Femoral, radial and large-bore alternative access.

Prog Cardiovasc Dis

January 2025

Charles R Drew University School of Medicine, Los Angeles, CA, United States of America; VA Long Beach Healthcare System, Long Beach, CA, United States of America. Electronic address:

Vascular access for coronary, peripheral, and structural interventions has changed significantly over the past two decades. The evolving demand for both large-bore access for valvular interventions and mechanical support devices, and for safer access for coronary interventions, in patients with comorbidities have driven progress in these areas. This review will provide an overview of the techniques of arterial access in the femoral, forearm (radial and ulnar), and alternative (transcarotid, transaxillary, and transcaval) locations based on the latest evidence and experience.

View Article and Find Full Text PDF

Comparison of single-catheter with two-catheter concept in older female patients undergoing transradial coronary angiography.

Herz

January 2025

Faculty of Medicine, Department of Cardiology, Bilecik Şeyh Edebali University, Pelitözü Mahallesi Fatih Sultan Mehmet Blv. No. 27 Merkez, Bilecik, Turkey.

Background: The aim of this study was to compare the efficacy outcomes of the traditional Judkins left and right catheters with those of the recently introduced Tiger catheter in female participants aged 65 years and older who underwent transradial coronary angiography.

Methods: A cohort of 160 female patients aged 65 and older who were scheduled for coronary angiography (CAG) were divided into two groups according to the use of Tiger (n = 80) or Judkins (n = 80) catheters for radial angiography at a ratio of 1:1, respectively. We analyzed the effectiveness of the Tiger and Judkins catheters, the incidence of catheter failure, contrast volume, CAG time, fluoroscopy time, dose-area product (in mGy/cm), and the occurrence of radial spasm.

View Article and Find Full Text PDF

Feasibility of Distal Radial Access in High Bleeding Risk Patients Who Underwent Percutaneous Coronary Intervention.

Korean Circ J

November 2024

Division of Cardiology, Department of Internal Medicine, Yonsei University College of Medicine and Cardiovascular Center, Yongin Severance Hospital, Yongin, Korea.

Backgrounds And Objectives: The distal radial access (DRA), a potential alternative to the trans-radial approach (TRA), may offer advantages in terms of access site complications due to its smaller vessel diameter, especially for high bleeding risk (HBR) patients. This study aims to investigate the feasibility of DRA in HBR patients.

Methods: Based on data from the KODRA registry, a prospective, multicenter cohort, this study analyzed 1,586 patients who underwent successful percutaneous coronary intervention (PCI) via DRA.

View Article and Find Full Text PDF