Rationale: Bioresorbable scaffolds may confer clinical benefit in long-term studies; early mechanistic studies using intravascular imaging have provided insightful information about the immediate and mid-term local serial effects of BRS on the coronary vessel wall.

Objectives: We assessed baseline, 6- and 12-month imaging data of the drug-eluting absorbable metal scaffold (DREAMS 2G).

Methods And Results: The international, first-in-man BIOSOLVE-II trial enrolled 123 patients with up to 2 de novo lesions (in vessels of 2.2 to 3.7mm). Angiographic based vasomotion, curvature and angulation were assessed; intravascular ultrasound (IVUS) derived radiofrequency (RF) data analysis and echogenicity were evaluated; optical coherence tomography (OCT) attenuation and backscattering analysis were also performed. There was hardly any difference in curvature between pre-procedure and 12months (-0.0019; p=0.48). The change in angulation from pre- to 12months was negligible (-3.58°; 95% CI [-5.97, -1.20]), but statistically significant. At 6months, the change in QCA based minimum lumen diameter in response to high dose of acetylcholine and IVUS-RF necrotic core percentage showed an inverse relationship (estimate of -0.489; p=0.055) and with fibrous volume a positive relationship (estimate of 0.53, p=0.035). Bioresorption analysis by OCT showed that the maximum attenuation values decreased significantly from post-procedure at 6months (Δ 6months vs. post-proc. is -13.5 [95% CI -14.6, -12.4]) and at 12months (Δ 12months vs. post-proc. is -14.0 [95% CI -15.4, -12.6]). By radiofrequency data, the percentage of dense calcium decreased significantly from post-procedure at 6months and at 12months. Likewise, by echogenicity, hyperechogenic structures decreased significantly from post-procedure at 6months; thereafter, they remained unchanged.

Conclusion: Following implantation of DREAMS 2G, restoration of the vessel geometry, vasomotion and bioresorption signs were observed at up to 12months; importantly, these changes occurred with preservation of the lumen size between 6 and 12months. NCT01960504.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijcard.2017.12.053DOI Listing

Publication Analysis

Top Keywords

decreased post-procedure
12
post-procedure 6months
12
drug-eluting absorbable
8
absorbable metal
8
metal scaffold
8
radiofrequency data
8
relationship estimate
8
12months
7
6months
5
vivo serial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!