In order for relaxation times to be used in clinical diagnosis, the precision of the measurement must be determined. The authors measured T1, T2, and proton density in a phantom and in human volunteers to determine the reproducibility of the method. The coefficient of variance of T1 measurements in the phantom during a 15-month period with two software upgrades was 5%. Variance of T2 measurements with any given software was 4% or less, and overall in the 15-month period, with two software changes, the T2 reproducibility was between 6% and 9%. The reproducibility is sufficiently high that precise clinical measurements of T1, T2, and proton density are feasible.