The utilization of hydrogen peroxide (H O ) cathodic reaction is an ideal approach to develop reliable biosensors that are immune to interferences arising from oxidizable endogenous/exogenous species in biological solutions. However, practical application of such a detection scheme is limited due to the significantly fluctuating oxygen levels in solutions, as oxygen can be reduced at similar potentials. Herein, this limitation is addressed by developing a novel electrode system with superhydrophobicity-mediated air-liquid-solid joint interfaces, which allows the rapid and continuous transport of oxygen from the air phase to the electrode surface and provides a fixed interfacial oxygen concentration. Using cathodic measurement of the enzymatic product H O , the sensing platform is applied to detect glucose, a model analyte, achieving a remarkably high selectivity (≈2% signal modulation due to common biologic interferents), sensitivity (18.56 µA cm mm ), and a dynamic linear range up to 80 × 10 m. The utility of H O reduction reaction at triphase interface to achieve reliable sensing platforms is general, and hence has broad potential in the fields of medical research, clinical diagnosis, and environmental analysis.