Measurement of effective detective quantum efficiency for a photon counting scanning mammography system and comparison with two flat panel full-field digital mammography systems.

Phys Med Biol

Radiation Physics Department, Queen's Centre for Oncology and Haematology, Castle Hill Hospital, Hull and East Yorkshire Hospitals NHS Trust, Castle Road, Hull, HU16 5JQ, United Kingdom. Faculty of Science, University of Hull, Cottingham Road, Hull, HU6 7RX, United Kingdom.

Published: January 2018

Effective detective quantum efficiency (eDQE) describes the resolution and noise properties of an imaging system along with scatter and primary transmission, all measured under clinically appropriate conditions. Effective dose efficiency (eDE) is the eDQE normalised to mean glandular dose and has been proposed as a useful metric for the optimisation of clinical imaging systems. The aim of this study was to develop a methodology for measuring eDQE and eDE on a Philips microdose mammography (MDM) L30 photon counting scanning system, and to compare performance with two conventional flat panel systems. A custom made lead-blocker was manufactured to enable the accurate determination of dose measurements, and modulation transfer functions were determined free-in-air at heights of 2, 4 and 6 cm above the breast support platform. eDQE were calculated for a Philips MDM L30, Hologic Dimensions and Siemens Inspiration digital mammography system for 2, 4 and 6 cm thick poly(methyl methacrylate) (PMMA). The beam qualities (target/filter and kilovoltage) assessed were those selected by the automatic exposure control, and anti-scatter grids were used where available. Measurements of eDQE demonstrate significant differences in performance between the slit- and scan-directions for the photon counting imaging system. MTF has been shown to be the limiting factor in the scan-direction, which results in a rapid fall in eDQE at mid-to-high spatial frequencies. A comparison with two flat panel mammography systems demonstrates that this may limit image quality for small details, such as micro-calcifications, which correlates with a more conventional image quality assessment with the CDMAM phantom. eDE has shown the scanning photon counting system offers superior performance for low spatial frequencies, which will be important for the detection of large low contrast masses. Both eDQE and eDE are proposed as useful metrics that should enable optimisation of the Philips MDM L30.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aaa307DOI Listing

Publication Analysis

Top Keywords

photon counting
16
flat panel
12
mdm l30
12
effective detective
8
detective quantum
8
quantum efficiency
8
counting scanning
8
mammography system
8
comparison flat
8
digital mammography
8

Similar Publications

Photon-counting mammography is an emerging modality that allows for spectral imaging and provides a differentiation of material compositions. The development of photon-counting mammography-specific contrast agents has yet to be explored. In this study, the contrast, sensitivity, and organ dose between silver sulfide nanoparticles (AgS-NPs) and a clinically approved iodinated agent (iopamidol) were investigated using a contrast-embedded gradient ramp phantom and a prototype scanner.

View Article and Find Full Text PDF

We describe a novel application of photon-counting detector CT (PCD-CT) in neurovascular imaging by harnessing the improved spatial resolution, attenuation of electronic noise, and reduction of metal artifacts. The presented case offers the unique challenge of high-quality imaging for the assessment of treated and untreated intracranial saccular aneurysms, in the setting of metal artifacts from embolization coils. Our goal was to explore optimized reconstruction parameters for ultra-high-resolution imaging (UHR) using a dedicated, sharp neurovascular kernel (Hv72) and the highest strength of quantum iterative reconstruction (QIR-4) for detailed characterization of the vasculature.

View Article and Find Full Text PDF

Background: To define optimal parameters for the evaluation of vessel visibility in intracranial stents (ICS) and flow diverters (FD) using photon-counting detector computed tomography angiography (PCD-CTA) with spectral reconstructions.

Methods: We retrospectively analyzed consecutive patients with implanted ICS or FD, who received a PCD-CTA between April 2023 and March 2024. Polyenergetic, virtual monoenergetic, pure lumen, and iodine reconstructions with different keV levels (40, 60, and 80) and reconstruction kernels (body vascular [Bv]48, Bv56, Bv64, Bv72, and Bv76) were evaluated by two radiologists with regions of interests and Likert scales.

View Article and Find Full Text PDF

Advanced Imaging of Type 2 Spinal CSF Leaks with Ultrahigh-Resolution Cone-Beam CT Myelography.

AJNR Am J Neuroradiol

January 2025

From Department of Neuroradiology (Niklas Lützen, Charlotte Zander, Horst Urbach), Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Germany and Department of Neurosurgery (Jürgen Beck, Florian Volz, Katharina Wolf, Amir El Rahal), Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Germany.

Type 2 CSF leaks are spinal lateral dural tears, causing spontaneous intracranial hypotension (SIH). They may be visualized with digital subtraction myelography (DSM), cone-beam CT (CBCT) myelography, energy-integrating detector or photon-counting CT myelography. A recently introduced ultrahigh-resolution cone-beam CT (UHR-CBCT) myelography has shown beneficial visualization of CSF-venous fistula, another cause of SIH.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!