Here, we discuss the case of a man with a history of ischemic cardiomyopathy and cardiac resynchronization therapy defibrillator implantation, who presented to emergency department with decompensated heart failure due to the loss of resynchronization therapy. The reason for the malfunction was left ventricle upper rate interval lock-in due to inappropriate programming of the device.