Background: Few studies have examined the use of continuous glucose monitoring (CGM) devices in the general wards. The aim of this pilot study was to examine whether CGM readings can be successfully transmitted from the bedside to a central monitoring device in the nursing station, and whether a glucose telemetry system can prevent hypoglycemic events.

Methods: We present pilot data on 5 consecutive insulin treated general medicine patients with type 2 diabetes (T2DM) whose glucose values were observed with CGM (DEXCOM) and the results were transmitted to a central nursing station monitoring system using DEXCOM Follow and Share 2 software. CGM alarms were set-up at glucose <85 mg/dl.

Results: Duration of CGM observation was 4.0 ± 1.6 days (mean ± SD). During CGM, the overall time spent within blood glucose (BG) target of 70-179 mg/dl was 64.68 ± 15% (mean ± SD), on hypoglycemia (<70 mg/dl) was 0.30% ± 0.39, and time spent on hyperglycemia (≥180 mg/dl) was 35.02% ± 15.5. Two patients had 3 actions of prevention of potential hypoglycemia (CGM BG <70 mg/dl for >20 minutes) captured by alarm. No patients had CGM glucose value <54 mg/dl.

Conclusions: This pilot study indicates that the use of CGM values in hospitalized patients can be successfully transmitted to a monitoring device in the nursing station, improving patient surveillance in insulin treated patients with diabetes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5761998PMC
http://dx.doi.org/10.1177/1932296817748964DOI Listing

Publication Analysis

Top Keywords

continuous glucose
8
glucose monitoring
8
general wards
8
glucose telemetry
8
telemetry system
8
nursing station
8
glucose
5
monitoring
4
monitoring preventing
4
preventing inpatient
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!