Quantitative assessment of reflux in commercially available needle-free IV connectors.

J Vasc Access

3 Department of Biomedical Engineering, University of Missouri, Columbia, MO - USA.

Published: January 2018

Introduction: Blood reflux is caused by changes in pressure within intravascular catheters upon connection or disconnection of a syringe or intravenous tubing from a needle-free connector (NFC). Changes in pressure, differing with each brand of NFC, may result in fluid movement and blood reflux that can contribute to intraluminal catheter occlusions and increase the potential for central-line associated bloodstream infections (CLABSI).

Methods: In this study, 14 NFC brands representing each of the four market-categories of NFCs were selected for evaluation of fluid movement occurring during connection and disconnection of a syringe. Study objectives were to 1) theoretically estimate amount of blood reflux volume in microliters (μL) permitted by each NFC based on exact component measurements, and 2) experimentally measure NFC volume of fluid movement for disconnection reflux of negative, neutral and anti-reflux NFC and fluid movement for connection reflux of positive displacement NFC.

Results: The results demonstrated fluid movement/reflux volumes of 9.73 μL to 50.34 μL for negative displacement, 3.60 μL to 10.80 μL for neutral displacement, and 0.02 μL to 1.73 μL for pressure-activated anti-reflux NFC. Separate experiment was performed measuring connection reflux of 18.23 μL to 38.83 μL for positive displacement NFC connectors.

Conclusions: This study revealed significant differences in reflux volumes for fluid displacement based on NFC design. While more research is needed on effects of blood reflux in catheters and NFCs, results highlight the need to consider NFCs based on performance of individual connector designs, rather than manufacturer designation of positive, negative and neutral marketing categories for NFCs without anti-reflux mechanisms.

Download full-text PDF

Source
http://dx.doi.org/10.5301/jva.5000781DOI Listing

Publication Analysis

Top Keywords

blood reflux
16
fluid movement
16
reflux
9
nfc
9
μl
9
changes pressure
8
connection disconnection
8
disconnection syringe
8
negative neutral
8
anti-reflux nfc
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!