Doppler ultrasound triggering for cardiac MRI at 7T.

Magn Reson Med

Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany.

Published: July 2018

Purpose: A prerequisite for cardiac MR (CMR) imaging is adequate synchronization of image acquisition with the cardiac cycle. Electrocardiogram triggering may be hampered by electromagnetic interferences at high field strength. The purpose of this work is to evaluate the feasibility of Doppler ultrasound triggering for CMR image synchronization at 7T ultra-high-field MRI.

Methods: A custom-built Doppler ultrasound (DUS) trigger device was developed. Magnetic resonance compatibility was evaluated using E- and H-field probes and flip angle maps prior to the study. Cardiac MR was performed at 7T in 13 healthy subjects using DUS and pulse oximetry for triggering. For validation of the trigger signal, the electrocardiogram, pulse, and DUS signals were compared outside of the MR room. Breath-hold cine fast low-angle-shot sequences were acquired in short-axis and four-chamber view. Image quality was assessed by two senior radiologists and by measurement of endocardial blurring.

Results: The maximal change in E- and H-field distributions with and without transducer was 5%. No interferences were observed between DUS and MRI in the B maps and during CMR imaging. Validation of the DUS trigger signal resulted in a high correlation to the electrocardiographic signal of r = 0.99. Analysis of image and trigger quality revealed no significant differences.

Conclusion: Doppler ultrasound was applied as a new trigger method in CMR at 7T. The transmission line and transducer were locally approved as 7T MR conditional, and were successfully tested for image synchronization at 7T. In the future, this method needs to be evaluated in a larger patient population. Magn Reson Med 80:239-247, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.27032DOI Listing

Publication Analysis

Top Keywords

doppler ultrasound
16
ultrasound triggering
8
cmr imaging
8
image synchronization
8
dus trigger
8
magnetic resonance
8
trigger signal
8
image
5
dus
5
trigger
5

Similar Publications

Color doppler blood flow analysis in circumscribed choroidal hemangioma.

Arq Bras Oftalmol

January 2025

Department of Ophthalmology & Visual Sciences, Escola Paulista de Medicina, Universidade Federal de São Paulo, São Paulo, SP, Brazil.

Purpose: To evaluate if color Doppler can detect internal blood flow in circumscribed choroidal hemangioma.

Methods: This cross-sectional study examined seven eyes of seven participants with circumscribed choroidal hemangiomas, with or without prior treatment. B-scan ultrasound and color Doppler were used to assess the dimensions, topographical distribution, and internal blood flow of the affected eyes.

View Article and Find Full Text PDF

The present study aims to examine the effect of 4 h of continuous sitting on cerebral endothelial function, which is a crucial component of cerebral blood flow regulation. We hypothesized that 4 h of sitting may impair cerebral endothelial function similarly to how it affects lower limb vasculature. Thirteen young, healthy participants were instructed to remain seated for 4 h without moving their lower limbs.

View Article and Find Full Text PDF

Optimizing the life of vascular access during follow-up.

J Cardiovasc Surg (Torino)

February 2025

Department of Vascular Surgery, ASST Settelaghi Universitary Teaching Hospital, University of Insubria, Varese, Italy.

Optimizing the longevity of vascular access in hemodialysis patients remains a critical aspect of patient care, given the significant role of arteriovenous fistulas (AVFs) and arteriovenous grafts (AVGs) in enabling effective dialysis. Vascular access complications, such as stenosis, thrombosis, and cannulation-related damage, continue to challenge both the functionality and the sustainability of these access points. Recent advancements underscore the importance of a robust follow-up strategy, integrating clinical evaluations with diagnostic tools like color Doppler ultrasound (CDU) and emerging interventional approaches such as drug-coated balloon (DCB) angioplasty.

View Article and Find Full Text PDF

Background: Testicular torsion (TT) is a urological emergency requiring prompt intervention to prevent irreversible damage to the testicle.

Objective: This study aims to assess trends in men's TT referrals, diagnostic evaluation through Doppler sonography (DS) scan, symptoms before surgery, orchidectomy rates, and TT laterality in relation to age and seasons of the year.

Methods: This observational retrospective cohort study included all patients treated for TT at King Abdullah University Hospital between 2009 and 2021.

View Article and Find Full Text PDF

Botulinum toxin-A (BTX-A) is widely used for aesthetic purposes and is recognized for its vasomotor actions. However, new medical applications have emerged. This study aims to describe the effect of BTX-A on human skin perfusion, particularly vessel diameter, blood flow, and blood vessel density in regions with known low vessel density.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!