A new approach for accurate mass assignment on a multi-turn time-of-flight mass spectrometer.

Eur J Mass Spectrom (Chichester)

2 Project Research Center for Fundamental Sciences, Graduate School of Science, Osaka University, Toyonaka, Japan.

Published: December 2017

A simple, effective accurate mass assignment procedure for a time-of-flight mass spectrometer is desirable. External mass calibration using a mass calibration standard together with an internal mass reference (lock mass) is a common technique for mass assignment, however, using polynomial fitting can result in mass-dependent errors. By using the multi-turn time-of-flight mass spectrometer infiTOF-UHV, we were able to obtain multiple time-of-flight data from an ion monitored under several different numbers of laps that was then used to calculate a mass calibration equation. We have developed a data acquisition system that simultaneously monitors spectra at several different lap conditions with on-the-fly centroid determination and scan law estimation, which is a function of acceleration voltage, flight path, and instrumental time delay. Less than 0.9 mDa mass errors were observed for assigned mass to charge ratios ( m/z) ranging between 4 and 134 using only Ar as a reference. It was also observed that estimating the scan law on-the-fly provides excellent mass drift compensation.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1469066717723755DOI Listing

Publication Analysis

Top Keywords

mass
14
mass assignment
12
time-of-flight mass
12
mass spectrometer
12
mass calibration
12
accurate mass
8
multi-turn time-of-flight
8
scan law
8
approach accurate
4
assignment multi-turn
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!