Background: 25-hydroxyvitamin D [25(OH)D] is the most reliable biomarker of vitamin D status, but until now 25(OH)D assays have suffered from inter-laboratory and inter-assay discrepancies. In the setting of the international Vitamin D Standardization Program, Immunodiagnostic Systems (IDS) recently reformulated and restandardized the 25(OH)D immunoassay available on the automated iSYS platform. In the present study, we evaluated this new generation of the 25(OH)D immunoassay (IS-2500).
Methods: Repeatability and within-laboratory imprecision were verified according to the Clinical and Laboratory Standards Institute EP15-A3. Results from the sera of 63 patients were compared with those obtained with the previous iSYS method (IS-2700S) using Passing-Bablok and Bland-Altman analysis. The prevalence and bias-adjusted kappa (PABAK) coefficient was calculated to assess the agreement of vitamin D status provided by the two iSYS immunoassays. Fourteen Vitamin D External Quality Assessment Scheme (DEQAS) samples were used to evaluate inaccuracy.
Results: Using the EP15-A3 protocol, repeatability and within-laboratory imprecision obtained with the new iSYS method were lower than 6% and 8%, respectively. These results are consistent with the manufacturer's claims. In more adverse conditions (50 measurements over 15days with multiple calibrations), the within-laboratory imprecision was 14.8% (39nmol/L) and 7.7% (155nmol/L). 25(OH)D concentrations measured with the new assay showed a strong correlation with those provided by the previous version (r=0.969, p<0.0001). The Passing-Bablok regression equation was as follows: new assay=1.079 x (previous assay) - 3.6nmol/L. The PABAK coefficient of 0.810 reflected almost perfect agreement between the two immunoassays to classify patients according to their vitamin D status (85.7% of agreement). Using DEQAS samples, the mean inaccuracy bias was lower than 5% when the new iSYS method was compared with LC-MS/MS methods and the NIST reference measurement procedure.
Conclusion: The new generation of the iSYS immunoassay evaluated in this study meets requirements for routinely measuring 25(OH)D levels in clinical laboratories.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiochem.2017.11.011 | DOI Listing |
Pract Lab Med
August 2024
Laboratory Dr. Limbach and Colleagues, Heidelberg, Germany.
Background: Clinical laboratories perform a wide range of tests that are used by healthcare professionals to guide medical decision making. Use of automated analyzers in the clinical laboratory can improve patient care by not only reducing the turn-around-time (TAT) of results but also improving accuracy of the reported results by reducing human error. The aim of this study was to evaluate the performance characteristics of a new automated laboratory instrument, the Atellica® CI Analyzer, Model 1900, over a 3-month period in a European laboratory setting.
View Article and Find Full Text PDFPract Lab Med
August 2024
Department of Clinical Laboratory, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Background: AMH is important in child growth and the concentrations change with age and gender. This study aimed to evaluate the performance of the Pylon AMH assays and establish pediatric reference intervals.
Methods: The experiments on imprecision, sensitivity, linearity, reportable range, interference and comparison were carried out to evaluate the analytical performance.
J Clin Lab Anal
August 2024
Department of Laboratory Medicine, Eunpyeong St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
Background: Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.
View Article and Find Full Text PDFClin Chem Lab Med
November 2024
20298 Hennepin Healthcare Research Institute, Minneapolis, MN, USA.
J Immunol Methods
July 2024
Institute for Clinical and Experimental Medicine, Department of Laboratory Methods, Vídeňská 1958/9, 140 21 Praha 4, Czech Republic; Third Faculty of Medicine, Charles University, Ruská 87, 100 00 Praha 10, Czech Republic.
Background: Manufacturers and diagnostic companies often recommend on-site verification of analytical performance in the clinical laboratory. The validation process used by manufacturers is rarely described in detail, and certain information on analytical performance is missing from the product sheet, especially for immunoanalytical methods. We describe an approach to the detailed validation of an ELISA method for the measurement of proprotein convertase subtilisin/kexin type 9 (PCSK9) plasma concentrations.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!