Evaluation of the new restandardized 25-hydroxyvitamin D assay on the iSYS platform.

Clin Biochem

Department of Biochemistry, University Hospital Centre Dijon-Burgundy, France; University Bourgogne Franche-Comté, LNC UMR1231, F-21000 Dijon, France; INSERM, LNC UMR1231, F-21000 Dijon, France.

Published: February 2018

Background: 25-hydroxyvitamin D [25(OH)D] is the most reliable biomarker of vitamin D status, but until now 25(OH)D assays have suffered from inter-laboratory and inter-assay discrepancies. In the setting of the international Vitamin D Standardization Program, Immunodiagnostic Systems (IDS) recently reformulated and restandardized the 25(OH)D immunoassay available on the automated iSYS platform. In the present study, we evaluated this new generation of the 25(OH)D immunoassay (IS-2500).

Methods: Repeatability and within-laboratory imprecision were verified according to the Clinical and Laboratory Standards Institute EP15-A3. Results from the sera of 63 patients were compared with those obtained with the previous iSYS method (IS-2700S) using Passing-Bablok and Bland-Altman analysis. The prevalence and bias-adjusted kappa (PABAK) coefficient was calculated to assess the agreement of vitamin D status provided by the two iSYS immunoassays. Fourteen Vitamin D External Quality Assessment Scheme (DEQAS) samples were used to evaluate inaccuracy.

Results: Using the EP15-A3 protocol, repeatability and within-laboratory imprecision obtained with the new iSYS method were lower than 6% and 8%, respectively. These results are consistent with the manufacturer's claims. In more adverse conditions (50 measurements over 15days with multiple calibrations), the within-laboratory imprecision was 14.8% (39nmol/L) and 7.7% (155nmol/L). 25(OH)D concentrations measured with the new assay showed a strong correlation with those provided by the previous version (r=0.969, p<0.0001). The Passing-Bablok regression equation was as follows: new assay=1.079 x (previous assay) - 3.6nmol/L. The PABAK coefficient of 0.810 reflected almost perfect agreement between the two immunoassays to classify patients according to their vitamin D status (85.7% of agreement). Using DEQAS samples, the mean inaccuracy bias was lower than 5% when the new iSYS method was compared with LC-MS/MS methods and the NIST reference measurement procedure.

Conclusion: The new generation of the iSYS immunoassay evaluated in this study meets requirements for routinely measuring 25(OH)D levels in clinical laboratories.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2017.11.011DOI Listing

Publication Analysis

Top Keywords

within-laboratory imprecision
12
isys platform
8
vitamin status
8
25ohd immunoassay
8
repeatability within-laboratory
8
isys method
8
isys
5
evaluation restandardized
4
restandardized 25-hydroxyvitamin
4
25-hydroxyvitamin assay
4

Similar Publications

Background: Clinical laboratories perform a wide range of tests that are used by healthcare professionals to guide medical decision making. Use of automated analyzers in the clinical laboratory can improve patient care by not only reducing the turn-around-time (TAT) of results but also improving accuracy of the reported results by reducing human error. The aim of this study was to evaluate the performance characteristics of a new automated laboratory instrument, the Atellica® CI Analyzer, Model 1900, over a 3-month period in a European laboratory setting.

View Article and Find Full Text PDF

Background: AMH is important in child growth and the concentrations change with age and gender. This study aimed to evaluate the performance of the Pylon AMH assays and establish pediatric reference intervals.

Methods: The experiments on imprecision, sensitivity, linearity, reportable range, interference and comparison were carried out to evaluate the analytical performance.

View Article and Find Full Text PDF

Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip.

J Clin Lab Anal

August 2024

Department of Laboratory Medicine, Eunpyeong St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.

Background: Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.

View Article and Find Full Text PDF
Article Synopsis
  • The study validated the Mindray high-sensitivity cardiac troponin I (hs-cTnI) assay by assessing its limits of blank (LoB) and detection (LoD), precision, linearity, analytical specificity, and sex-specific reference limits.
  • Key findings showed that the assay meets the standards of a high-sensitivity test, with a LoB and LoD both under 0.1 ng/L, good precision (1.2-3.8% variation), and acceptable specificity for various interfering substances.
  • The 99th percentile upper reference limits for cardiac troponin were determined to be 10 ng/L overall, with 5 ng/L for females and 12 ng/L for males, confirming the
View Article and Find Full Text PDF

How to verify the analytical and clinical performance of ELISA immunoanalysis in the real laboratory practice. PCSK9 as an example.

J Immunol Methods

July 2024

Institute for Clinical and Experimental Medicine, Department of Laboratory Methods, Vídeňská 1958/9, 140 21 Praha 4, Czech Republic; Third Faculty of Medicine, Charles University, Ruská 87, 100 00 Praha 10, Czech Republic.

Background: Manufacturers and diagnostic companies often recommend on-site verification of analytical performance in the clinical laboratory. The validation process used by manufacturers is rarely described in detail, and certain information on analytical performance is missing from the product sheet, especially for immunoanalytical methods. We describe an approach to the detailed validation of an ELISA method for the measurement of proprotein convertase subtilisin/kexin type 9 (PCSK9) plasma concentrations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!