The widely-used gamma-emitter Tc-99m has been shown to lead to optical emissions in mice and glass. We investigated the possibility that these emissions are due to the Cerenkov effect and whether the light emitted is proportional to local dose. By using a Geant4 Monte Carlo model matched to an experimental measurement, we show that the light detected by a small animal optical imaging system provides a 2D map of the dose throughout a glass sample. We conclude that radioluminescence from Tc-99m can be used to quantitatively measure dose in transparent materials, which could have applications in dosimetry and quality assurance.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2017.09.119 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!