Aim: To evaluate the safety and efficacy of the M4 (porous polyethylene plate) Ahmed Glaucoma Valve (AGV) drainage implant in a multicenter retrospective study.

Materials And Methods: A retrospective chart review of medical records of patients who had undergone the M4 Ahmed valve was performed from January 2013 to April 2015. The primary outcome measure was surgical failure defined as: Less than a 20% reduction in baseline intraocular pressure (IOP) to last follow-up visit, final IOP less than 5 mm Hg or greater than 18 mm Hg, reoperation for glaucoma, or loss of light perception vision. All eyes not meeting the above criteria were defined as success.

Results: A total of 291 eyes met all study inclusion criteria. The average follow-up in the study was 6 months (±7.6 months) with 112 patients achieving 12-month follow-up (38.5%). 208 eyes (71.5%) met the study success criteria at final follow-up. No statistically significant spikes in postoperative IOP at 1 and 4 months were detected. The average preoperative IOP was 26.0 on an average of 2.8 medications. At 6 months, the average IOP dropped to 16.7 on 0.9 medications and stayed relatively stable at 15.8 on 1.2 medications at 12-month follow-up.

Conclusion: The M4 valve appears to have less of a hypertensive phase compared with the other Ahmed class valves with a similar safety profile. While 71.5% success rate was achieved at final follow-up, the failure rate steadily increased over time.

Clinical Significance: While the M4 production has been discontinued, the porous design of the M4 may avoid a pressure spike in the Ahmed valve class and warrants future investigation for valve design. Sluch I, Gudgel B, Dvorak J, Ahluwalia MA, Ding K, Vold S, Sarkisian S. Clinical Experience with the M4 Ahmed Glaucoma Drainage Implant. J Curr Glaucoma Pract 2017;11(3):92-96.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5684239PMC
http://dx.doi.org/10.5005/jp-journals-10028-1231DOI Listing

Publication Analysis

Top Keywords

ahmed glaucoma
12
drainage implant
12
clinical experience
8
experience ahmed
8
glaucoma drainage
8
ahmed valve
8
met study
8
final follow-up
8
ahmed
6
glaucoma
5

Similar Publications

Importance: While urban counties maintain higher densities of ophthalmologists than rural counties, the geographic distribution of ophthalmic surgical subspecialists has not yet been elucidated. A potential workforce discrepancy may impact the burden of care faced by rural surgeons.

Objective: To assess the geographic distribution of the ophthalmic subspecialist surgeon workforce and evaluate factors associated with practicing in rural areas.

View Article and Find Full Text PDF

Altwijri and Alsirhy reported a case of uveitis-glaucoma-hyphema syndrome after an Ahmed glaucoma valve implantation surgery in an advanced primary open-angle glaucoma patient, being the first ever recorded of its kind. The author describes the position of the tube as the origin of the anterior chamber inflammation and hyphema, which resolved shortly after shortening and relocating it. This publication emphasizes the importance of precise implant positioning and close-up patient follow-up after glaucoma filtration surgery as an important standard for healthcare providers.

View Article and Find Full Text PDF

Purpose: To report cytokine/chemokine profiles of ocular fluid in two patients with herpetic uveitis.

Methods: Cytokine and chemokine profiling of ocular fluid was performed in two patients with herpetic uveitis. Ocular fluid findings were correlated with disease manifestations and the patients' clinical course.

View Article and Find Full Text PDF

Purpose: To evaluate the efficacy and safety of Ahmed glaucoma valve (AGV) implantation with subsequent trans-scleral diode cyclophotocoagulation (CPC) as the main intervention if IOP remained medically uncontrolled.

Patients And Methods: Charts of 108 consecutive eyes (90 patients) that underwent AGV implantation from 2003 to 2018 at a single clinical practice were retrospectively reviewed. The procedure was considered a failure if any of the following occurred: additional incisional glaucoma surgery, IOP >21 mmHg or < 20% reduction from baseline on 2 consecutive study visits after 3 months, IOP ≤ 5 mmHg on 2 consecutive study visits after 3 months, loss of light perception, or AGV removal.

View Article and Find Full Text PDF

An 18-year-old woman with a congenital iris stromal cyst was referred for sudden-onset increased intraocular pressure (IOP) that did not respond to medical therapy. Ultrasound biomicroscopy demonstrated synechial angle closure from 9:00 to 12:00 in addition to a large cyst occupying much of the inferior angle. After implantation of an Ahmed glaucoma valve, her visual acuity returned to baseline, and IOP normalized.

View Article and Find Full Text PDF