Application of the sum-peak method to activity standardizations of Eu sources in LNMRI (BR).

Appl Radiat Isot

Laboratório de Instrumentação Nuclear (LIN/PEN/COPPE/UFRJ), Caixa Postal 68509, CEP 21945-970 Rio de Janeiro, RJ, Brazil. Electronic address:

Published: April 2018

The sum-peak method, an absolute measurement technique that uses coincidence counting and γ-ray spectrometry, was applied to activity standardization of Eu sources. The decay branch used was that of Sm, as it is almost entirely electron capture and exhibits X-rays with a high probability of coincidence with the 121.8keVgamma rays, as evidenced by the matrix technique used for complex decay schemes, a method for deriving counting rate equations describing coincidence summing of gamma and X-rays. The result was compared with the calibration performed by ionization chamber traceable to BIPM. The results presented uncertainty values of 0.50% (k = 1).