Purpose: To find out whether 30-2 Swedish Interactive Threshold Algorithm (SITA) Fast is comparable to 30-2 SITA Standard as a tool for perimetry among the patients with intracranial tumors.

Methods: This was a prospective cross-sectional study involving 80 patients aged ≥18 years with imaging proven intracranial tumors and visual acuity better than 20/60. The patients underwent multiple visual field examinations using the two algorithms till consistent and repeatable results were obtained.

Results: A total of 140 eyes of 80 patients were analyzed. Almost 60% of patients undergoing perimetry with SITA Standard required two or more sessions to obtain consistent results, whereas the same could be obtained in 81.42% with SITA Fast in the first session itself. Of 140 eyes, 70 eyes had recordable field defects and the rest had no defects as detected by either of the two algorithms. Mean deviation (MD) (P = 0.56), pattern standard deviation (PSD) (P = 0.22), visual field index (P = 0.83) and number of depressed points at P < 5%, 2%, 1%, and 0.5% on MD and PSD probability plots showed no statistically significant difference between two algorithms. Bland-Altman test showed that considerable variability existed between two algorithms.

Conclusion: Perimetry performed by SITA Standard and SITA Fast algorithm of Humphrey Field Analyzer gives comparable results among the patients of intracranial tumors. Being more time efficient and with a shorter learning curve, SITA Fast my be recommended as a standard test for the purpose of perimetry among these patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5700593PMC
http://dx.doi.org/10.4103/ijo.IJO_315_17DOI Listing

Publication Analysis

Top Keywords

sita fast
16
perimetry patients
12
patients intracranial
12
intracranial tumors
12
sita standard
12
swedish interactive
8
interactive threshold
8
threshold algorithm
8
algorithm humphrey
8
humphrey field
8

Similar Publications

Purpose: This study compared the agreement between the Humphrey Field Analyzer (HFA) SITA Fast strategy and a novel virtual reality head-mounted visual perimetry device (VisuALL) in glaucoma patients.

Design: This is prospective observational study.

Participants: This study was conducted on 62 eyes of 39 glaucoma subjects.

View Article and Find Full Text PDF

Purpose: This study evaluates differences in the visual field performance when wearing the Defocus Incorporated Multiple Segments (DIMS) spectacle lens compared to wearing a conventional single vision (SV) spectacle lens.

Methods: Twenty-one children aged 9-14 years with spherical equivalent refraction (SER) between -1.13D to -4.

View Article and Find Full Text PDF

Introduction: A slower adoption rate of fixed dose combinations (FDC) in diabetes management is partly due to insufficient data. This study evaluates the safety and efficacy of an FDC of dapagliflozin + sitagliptin + metformin hydrochloride extended release (XR), compared to a dual FDC of sitagliptin + metformin hydrochloride XR among patients with type 2 diabetes mellitus (T2DM) with poor glycemic control when treated with metformin monotherapy.

Methods: A total of 274 patients with T2DM were randomized (1:1) to either arm X, receiving FDC of dapagliflozin (10 mg) + sitagliptin (100 mg) + metformin hydrochloride XR (1000 mg) (Dapa + Sita + Met) tablets, or arm Y, receiving sitagliptin phosphate (100 mg) + metformin hydrochloride XR (1000 mg) (Sita + Met) tablets, and treated for 16 weeks.

View Article and Find Full Text PDF

Optical Coherence Tomography Abnormalities as the Presenting Sign of an Involuted Sellar/Suprasellar Mass.

Case Rep Ophthalmol

October 2024

Department of Ophthalmology and Vision Sciences, Department of Medicine, University of Toronto, Toronto, ON, Canada.

Introduction: Pituitary adenomas are benign tumours that can lead to visual loss through compression of the optic chiasm. Patients with pituitary adenomas often present with visual field defects (commonly bitemporal hemianopia), but some may be asymptomatic. In such cases, abnormalities may only be detected through visual field testing or optical coherence tomography (OCT) of the ganglion cell-inner plexiform layer (GCIPL), which may provide a more sensitive method for detecting such abnormalities.

View Article and Find Full Text PDF

Purpose: To compare two fast threshold strategies of visual field assessment: SITA-Fast (SF; Humphrey field analyser) and Elisar-Fast (EF; Advanced vision analyser) in patients with glaucoma.

Methods: In this cross-sectional observational study, of total 192 subjects, 138 subjects [150 eyes, 80 glaucoma subjects (91 eyes) and 58 healthy controls (59 eyes)] were analysed and included. Each subject underwent 24-2 EF and SF in randomized order with a minimum time interval of 1 hour between tests.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!