Purpose: To evaluate the effect on charge collection in the ionization chamber (IC) in proton pencil beam scanning (PBS), where the local dose rate may exceed the dose rates encountered in conventional MV therapy by up to three orders of magnitude.

Methods: We measured values of the ion recombination (k ) and polarity (k ) correction factors in water, for a plane-parallel Markus TM23343 IC, using the cyclotron-based Proteus-235 therapy system with an active proton PBS of energies 30-230 MeV. Values of k were determined from extrapolation of the saturation curve and the Two-Voltage Method (TVM), for planar fields. We compared our experimental results with those obtained from theoretical calculations. The PBS dose rates were estimated by combining direct IC measurements with results of simulations performed using the FLUKA MC code. Values of k were also determined by the TVM for uniformly irradiated volumes over different ranges and modulation depths of the proton PBS, with or without range shifter.

Results: By measuring charge collection efficiency versus applied IC voltage, we confirmed that, with respect to ion recombination, our proton PBS represents a continuous beam. For a given chamber parameter, e.g., nominal voltage, the value of k depends on the energy and the dose rate of the proton PBS, reaching c. 0.5% for the TVM, at the dose rate of 13.4 Gy/s. For uniformly irradiated regular volumes, the k value was significantly smaller, within 0.2% or 0.3% for irradiations with or without range shifter, respectively. Within measurement uncertainty, the average value of k , for the Markus TM23343 IC, was close to unity over the whole investigated range of clinical proton beam energies.

Conclusion: While no polarity effect was observed for the Markus TM23343 IC in our pencil scanning proton beam system, the effect of volume recombination cannot be ignored.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.12668DOI Listing

Publication Analysis

Top Keywords

proton pbs
16
ion recombination
12
dose rate
12
markus tm23343
12
recombination polarity
8
polarity correction
8
correction factors
8
ionization chamber
8
proton
8
chamber proton
8

Similar Publications

Tenapanor: A novel therapeutic agent for dialysis patients with hyperphosphatemia.

Ther Apher Dial

January 2025

Division of Nephrology, Endocrinology, and Metabolism, Department of Internal Medicine, Tokai University School of Medicine, Isehara, Kanagawa, Japan.

Patients on dialysis often develop hyperphosphatemia, contributing to an increased risk of cardiovascular events and mortality. Currently, several types of phosphate binders (PBs) exist for the treatment of hyperphosphatemia, but they are sometimes associated with drug-specific side effects and high pill burden, making it difficult to control serum phosphorus appropriately. Tenapanor, which has a novel mechanism to reduce serum phosphorus via selective sodium/proton exchange transporter 3 inhibition, was approved for hyperphosphatemia in Japan in 2023.

View Article and Find Full Text PDF

Background And Purpose: Radiation induced image changes (IC) on MRI have been observed after proton therapy for brain tumours. This study aims to create predictive models, with and without taking into account patient variation, based on dose, linear energy transfer (LET) and periventricular zone (PVZ) in a national cohort of patients with glioma treated with pencil beam scanning (PBS).

Materials And Methods: A cohort of 87 consecutive patients with oligodendroglioma or astrocytoma (WHO grade 2-4) treated with PBS from January 2019 to December 2021 was included.

View Article and Find Full Text PDF

Purpose: This study evaluates the hypothesis that a volumetric skin-sparing planning technique (SSPT) will reduce acute dermatitis in patients treated to the breast or chest wall (CW) with proton pencil-beam scanning (PBS).

Methods And Materials: In January 2022, our center incorporated volumetric-based skin-sparing objectives in addition to skin hot spot evaluation as an SSPT. The SSPT incorporated an objective to limit the volume of a skin evaluation structure (skin-eval) receiving 95% of the prescription dose or more (V95%Rx) to ideally < 50%.

View Article and Find Full Text PDF

pH sensing technology is pivotal for monitoring aquatic ecosystems and diagnosing human health conditions. Indium-gallium-zinc oxide electrolyte-gated thin-film transistors (IGZO EGTFTs) are highly regarded as ion-sensing devices due to the pH-dependent surface chemistry of their sensing membranes. However, applying EGTFT-based pH sensors in complex biofluids containing diverse charged species poses challenges due to ion interference and inherently low sensitivity constrained by the Nernst limit.

View Article and Find Full Text PDF

An AI dose-influence matrix engine for robust pencil beam scanning protons therapy.

Med Phys

December 2024

National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, People's Republic of China.

Background: Rapid planning is of tremendous value in proton pencil beam scanning (PBS) therapy in overcoming range uncertainty. However, the dose calculation of the dose influence matrix (D) in robust PBS plan optimization is time-consuming and requires substantial acceleration to enhance efficiency.

Purpose: To accelerate the D calculations in PBS therapy, we developed an AI-D engine integrated into our in-house treatment planning system (TPS).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!