Background: This study evaluated the clinical efficacy of a novel imaging system (HyperEye Medical System [HEMS]; Mizuho Corp., Tokyo, Japan) that uses the near-infrared (NIR) fluorescence of indocyanine green to analyze sentinel lymph node (SLN) biopsies for the staging of breast cancer.

Methods: This study enrolled 91 patients with histologically confirmed breast cancer that was clinically node negative with a tumor size <3 cm. We compared SLN identification rates between HEMS and conventional methods (gamma probe scanning using a colloidal radioisotope [RI] and a blue dye method) by analyzing the relationships of lymphatic to axillary lesions and SLNs.

Results: The identification rate of SLNs was 100% using HEMS, 97.8% using the RI method, and 95.6% using the blue dye method. Two types of lymphatic pathway (LP) were detected in 39 patients (42.9%) and also clearly identified using HEMS-captured color and NIR fluorescence. The incidence of two or more SLNs was significantly higher in patients with a two-route LP to the axilla group than in those with only one route (p < 0.001; 43.6 vs. 9.6%).

Conclusions: The HEMS NIR fluorescence color imaging method is a promising potential modality for higher-level identification of SLNs than a standard combination of the RI and blue dye methods.

Download full-text PDF

Source
http://dx.doi.org/10.1159/000484050DOI Listing

Publication Analysis

Top Keywords

sentinel lymph
8
breast cancer
8
novel color
4
color fluorescence
4
fluorescence navigation
4
navigation system
4
system intraoperative
4
intraoperative transcutaneous
4
transcutaneous lymphatic
4
lymphatic mapping
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!