Development and validation of a stability-indicating HPLC-UV method for the determination of triamcinolone acetonide and its degradation products in an ointment formulation.

J Pharm Biomed Anal

Department of Pharmaceutics, Utrecht Institute for Pharmaceutical Sciences, Utrecht University, 3584 CG Utrecht, The Netherlands; Research and Development Department, Tiofarma B.V., Hermanus Boerhaavestraat 1, 3261 ME Oud-Beijerland, The Netherlands; Department of Clinical Pharmacy, Division of Laboratory Medicine & Pharmacy, University Medical Centre Utrecht, P/O Box 85500, 3508 GA Utrecht, The Netherlands.

Published: February 2018

A stability indicating high performance liquid chromatography method has been developed for the determination of triamcinolone acetonide (TCA) and its main degradation products in ointment formulations. The method, based on extensive stress testing using metal salts, azobisisobutyronitrile, acid, base and peroxide, showed that TCA undergoes oxidative degradation. All degradation products were identified using HPLC mass spectrometry. Separation and quantification was achieved using an Altima C18 RP18 HP column (250×4.6mm, with 5μm particles) using a mobile phase consisting of acetonitrile and water buffered at pH 7 using 10mM phosphate buffer. A gradient mode was operated at a flow rate of 1.5ml/min and detection was at 241nm. The method showed linearity for TCA and Impurity C in 0.02-125% of the workload, both square roots of the correlation coefficients were larger than 0.9999. Repeatability and intermediate precision were performed by six consecutive injections of both 1.25% and 125% of the work load for both TCA and Impurity C divided equally over two days. RSD were 0.6% and 0.7% for TCA and 0.5% and 0.1% for Impurity C respectively. Accuracy was determined as well, the average recoveries were 99.5% (±0.1%, n=3) for TCA and 96.9% (±1.3%, n=3) for impurity C respectively from spiked ointment samples. The robustness was also evaluated by variations of column (old vs new), mobile phase pH and filter retention. The applicability of the method was evaluated by analysis of a commercial ointment formulation. Interestingly, the extensive stress tests were able to predict all degradation products of TCA in a long term stability ointment sample.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jpba.2017.11.026DOI Listing

Publication Analysis

Top Keywords

degradation products
16
determination triamcinolone
8
triamcinolone acetonide
8
products ointment
8
ointment formulation
8
extensive stress
8
mobile phase
8
tca impurity
8
tca
7
method
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!