Background: Glycopyrrolate administered by a novel, investigational eFlow Closed System (CS) nebulizer (eFlow CS) is being evaluated for the maintenance treatment of chronic obstructive pulmonary disease (COPD). The eFlow CS is a hand-held, vibrating membrane nebulizer optimized to deliver 1 mL of glycopyrrolate solution into the lung in <3 minutes. Clinical studies have shown improvements in lung function of subjects treated with nebulized glycopyrrolate.

Methods: The aerosol performance of the eFlow CS nebulizer was characterized by delivered dose, aerodynamic droplet size distribution and nebulization time. Simulated use nebulizer performance over 60 days was assessed by volume median diameter (VMD), nebulized amount, and nebulization time. Nebulization outputs were assayed to ensure adequate delivery of glycopyrrolate with an acceptable impurity profile. Aerosol condensates were analyzed for glycopyrrolate concentration and impurities by ultra-high-performance liquid chromatography and compared with non-nebulized samples.

Results: The mean mass median aerodynamic diameter, geometric standard deviation, and fine particle fraction were 3.7 μm, 1.7, and 72%, respectively, and independent of formulation strength (25 and 50 μg/mL). Delivered dose was 88% of the nominal dose for both formulation strengths. The mean delivered dose, assessed by breathing simulation, was 56.8% for 25 μg/mL and 62.6% for 50 μg/mL. Nebulization times were 1-2.5 minutes with no apparent increasing trend with use over a 60-day period. The nebulized amount showed no significant changes, whereas the VMD showed a slight, but not pharmaceutically relevant, increase (0.1-0.2 μm) after 60-day simulated use. Glycopyrrolate concentration and impurity levels of nebulized samples were statistically similar to those of non-nebulized samples.

Conclusion: The eFlow CS generates glycopyrrolate aerosols with high delivered dose, short treatment time, and small droplet size with narrow size distribution suitable for central and peripheral airway deposition. The unit dose vial mitigates medication misuse and ensures dose uniformity. Results support the use of glycopyrrolate/eFlow CS for the treatment of COPD.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5994673PMC
http://dx.doi.org/10.1089/jamp.2017.1384DOI Listing

Publication Analysis

Top Keywords

eflow closed
8
closed system
8
system nebulizer
8
vitro characterization
4
eflow
4
characterization eflow
4
nebulizer glycopyrrolate
4
glycopyrrolate inhalation
4
inhalation solution
4
solution background
4

Similar Publications

Recent research has greatly focused on the environmental water supplement of rivers individually and independently. However, a comprehensive and integrated view of all rivers in the basin is simultaneously required in closed basins leading to lakes and wetlands. This has affected Lake Urmia, which is the second largest saltwater lake in the world.

View Article and Find Full Text PDF

Inhaled bronchodilators are central for the treatment of chronic obstructive pulmonary disease (COPD), as they can provide symptom relief and reduce the frequency and severity of exacerbations while improving health status and exercise tolerance. In 2017, glycopyrrolate (GLY) delivered via the eFlow® closed system (CS) nebulizer (nebulized GLY; 25 µg twice daily), was approved by the US Food and Drug Administration for maintenance treatment of moderate-to-very-severe COPD. This approval was based largely on results from the replicate, placebo-controlled, Phase III clinical trials- GOLDEN 3 and 4.

View Article and Find Full Text PDF

Robustness of aerosol delivery of amikacin liposome inhalation suspension using the eFlow® Technology.

Eur J Pharm Biopharm

September 2021

Insmed Incorporated, 700 US 206 North, Bridgewater, NJ 08807, USA. Electronic address:

The purpose of these studies was to understand the effect on product performance of batch-to-batch variability in both the amikacin liposome inhalation suspension (ALIS) formulation and its delivery device, the Lamira® nebulizer system, designed and manufactured by PARI (PARI Pharma GmbH, Munich, Germany). Three batches of ALIS spanning a range of lipid concentrations (43, 48 and 54 mg/mL) were tested with nine PARI inhalation devices that varied within the production process of the vibrating membrane with respect to hole geometry. Three hole geometry clusters were built including a geometry close to the mean geometry (median) and two geometries deviating from the mean geometry with smaller (smaller) and larger (larger) holes.

View Article and Find Full Text PDF

Aims: This study compared medication use, healthcare resource utilization (HRU), and exacerbations among individuals with chronic obstructive pulmonary disease (COPD) who initiated glycopyrrolate/eFlow Closed System nebulizer 25 mcg/mL glycopyrrolate (hereafter GLY) in a real-world setting before and after treatment initiation.

Materials And Methods: Retrospective claims and hospital charge master data were used to identify individuals ≥ 40 years of age diagnosed with COPD who initiated GLY between 1 April 2018 and 28 February 2019 (first prescription claim = index date). Patients were excluded if they had ≥1 asthma diagnosis in the 6-month pre-index period.

View Article and Find Full Text PDF
Article Synopsis
  • The study analyzed data from two Phase III trials to see if glycopyrrolate inhalation (GLY) helps reduce the risk of clinically important deterioration (CID) in COPD patients.
  • GLY at doses of 25 mcg and 50 mcg taken twice daily significantly decreased the risk of CID by 50% and 40%, respectively, compared to placebo.
  • The treatment was particularly effective among younger patients and those with lower peak inspiratory flow rates, demonstrating that GLY can be beneficial across various demographics and COPD severities.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!