Electrocardiographic patterns in biventricular pacing delivered by second-generation cardiac resynchronization devices.

Indian Pacing Electrophysiol J

Cardiac Electrophysiology Research Center, Rajaie Cardiovascular Medical and Research Center, Iran University of Medical Sciences, Tehran, Iran; Department of Cardiac Electrophysiology, Rajaie Cardiovascular Medical and Research Center, Iran University of Medical Sciences, Tehran, Iran. Electronic address:

Published: November 2017

Background: With increasing use of cardiac resynchronization therapy (CRT), treating physicians should be familiar with different electrocardiographic (ECG) patterns of left ventricular (LV) lead and biventricular (BiV) pacing. However, there are a few publications on ECG patterns during BiV pacing.

Purpose: This study was sought to determine different ECG patterns in patients with BiV pacing.

Methods: Twelve-lead ECGs during BiV pacing (right ventricular leads at apex and LV leads in one of the lateral coronary veins) were analyzed in 181 consecutive patients (121 male; mean age, 62.0 ± 13.5 years) with advanced heart failure and baseline left bundle branch block pattern after at least 6-month of uncomplicated CRT.

Results: During BiV pacing, 65% of the patients showed a dominant R wave in V1. There was a right axis deviation in 57% in frontal plane. However, a left superior axis emerged in 34% and normal frontal plane axis in 9%. Sequential BiV pacing (73% vs. 58%, P = 0.04) and pacing from posterolateral coronary vein (80% vs. 60%, p = 0.045) were more likely to present with a dominant R wave in V1. In sequential pacing, AV interval was significantly longer in patients with negative complex in V1 than in those with positive complex (124 ± 21 vs. 116 ± 8.0, p = 0.005). A Q/q wave was detected in 85% of patients in lead I and 78% in lead aVL.

Conclusions: BiV pacing from lateral coronary venous branches and right ventricular apex characteristically presented with dominant R wave in V1, Q/q wave in leads I and aVL, and right or left superior axis. However, a negative complex in V1, QRS axis in other quadrants, and lack of Q/q wave in leads I and aVL did not necessarily indicate a problem.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5840759PMC
http://dx.doi.org/10.1016/j.ipej.2017.10.007DOI Listing

Publication Analysis

Top Keywords

biv pacing
20
ecg patterns
12
dominant wave
12
q/q wave
12
pacing
8
cardiac resynchronization
8
lateral coronary
8
frontal plane
8
left superior
8
superior axis
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!