Prism under cover test in alternate fixation horizontal strabismus.

Curr Eye Res

a Department of Ophthalmology , Hospital Ramón y Cajal, Madrid , Spain.

Published: February 2018

Purpose: To evaluate the applicability of the prism under cover test (PUCT) to quantify manifest deviation in horizontal strabismus with alternate fixation when simultaneous prism and cover test (SPCT) is not feasible.

Methods: Children aged 4-11 years, with alternate fixation horizontal strabismus and alternate prism and cover test (APCT) distance deviation (DD) up to 25 PD were eligible. In group 1 of the study, SPCT was not feasible (n = 18), whereas in group 2, it was feasible (n = 24). Refraction, PUCT, APCT, central/peripheral fusion, and stereoacuity were measured. Repeatability of PUCT, agreement between PUCT and SPCT, and Pearson correlations between variables were studied.

Results: In group 1, mean DD was 15.6 and 9.5 PD using APCT and PUCT, respectively (intraclass correlation, ICC: 0.90). Mean stereoacuity was 201.1 arc seconds. PUCT was better correlated with stereoacuity than APCT, but not significantly (p = 0.12). Coefficient of repeatability for PUCT was 3.4 PD. In group 2, mean DD was 17.6, 14.1, and 12.5 PD using APCT, PUCT, and SPCT, respectively. Mean stereoacuity was 285.9 arc seconds. Correlations APCT-PUCT (0.87), APCT-SPCT (0.82), and SPCT-PUCT (0.95) were significant. APCT did not show as good correlation with stereoacuity (0.58) as SPCT and PUCT (0.74 and 0.78, respectively). Concordance correlation coefficient between SPCT and PUCT was 0.91, and Bland Altman agreement between the two variables was also good.

Conclusions: PUCT is a procedure with good repeatability, of interest in children to estimate manifest deviation when SPCT is not feasible, and in support of the diagnosis of monofixation syndrome.

Download full-text PDF

Source
http://dx.doi.org/10.1080/02713683.2017.1385085DOI Listing

Publication Analysis

Top Keywords

prism cover
16
cover test
16
alternate fixation
12
horizontal strabismus
12
puct
11
fixation horizontal
8
manifest deviation
8
strabismus alternate
8
spct feasible
8
repeatability puct
8

Similar Publications

Purpose: To use swept-source optical coherence tomography angiography (SS-OCTA) to investigate the alterations in retinal vascular density (VD) in patients presenting with congenital unilateral trochlear nerve palsy.

Methods: The medical records of patients diagnosed with congenital unilateral trochlear nerve palsy and those of a healthy control group were reviewed retrospectively. Comprehensive ocular examinations and SS-OCTA imaging were conducted.

View Article and Find Full Text PDF

Background: Colorectal cancer (CRC) is one of thes most prevalent malignant tumors worldwide. Current therapeutic strategies for CRC have limitations, while nanomaterials show significant potential for diagnosing and treating CRC. This study utilizes bibliometric analysis to evaluate the current status and trends in this field.

View Article and Find Full Text PDF

Ultra-light antennas via charge programmed deposition additive manufacturing.

Nat Commun

January 2025

Advanced Manufacturing and Metamaterials Laboratory, Department of Material Science and Engineering, University of California, Berkeley, CA, USA.

The demand for lightweight antennas in 5 G/6 G communication, wearables, and aerospace applications is rapidly growing. However, standard manufacturing techniques are limited in structural complexity and easy integration of multiple material classes. Here we introduce charge programmed multi-material additive manufacturing platform, offering unparalleled flexibility in antenna design and the capability for rapid printing of intricate antenna structures that are unprecedented or necessitate a series of fabrication routes.

View Article and Find Full Text PDF

Current clinical measurements of strabismus angles (SAs) are manual and require a skilled examiner and active cooperation by patients. This results in high interexaminer variability, and clinical use is limited by the availability of trained examiners. An objective and automated procedure, independent of the examiner, would be useful.

View Article and Find Full Text PDF

Accuracy of virtual reality in automated measurement and diagnosis of strabismus.

Digit Health

December 2024

Beijing Tongren Eye Center, Beijing Tongren Hospital, Beijing Key Laboratory of Ophthalmology & Visual Sciences, Capital Medical University, Beijing, China.

Background: To explore the accuracy of VR technology for strabismus screening.

Methods: In this cross-sectional study, a total of 191 subjects were included. Examinations included VR technology based on infrared, the cover-uncover test, and the prism alternating cover test (PACT) at various gaze positions, specifically at distances of 33 cm and 6 m.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!