Given the limited access to healthcare resources, low-income settings require the development of affordable technology. Here we present the design and evaluation of a low-cost colorimeter applied to the non-invasive monitoring of through the detection of glucose in salival fluid. Samples were processed by the glucose oxidase-peroxidase enzymatic system and analyzed with the development equipment. A light emission diode of 532.5 nm was used as an excitation source and a RGB module was used as a receptor. A calibration curve to quantify the concentration of salivary glucose (0 to 18 mg/dL) was carried out by relating the RGB components registered with glucose concentrations, achieving a limit of detection of 0.17 mg/dL with a CV of 5% (n = 3). Salivary samples of diabetic and healthy volunteers were processed with the equipment showing an average concentration of 1.5519 ± 0.4511 mg/dL for the first and 4.0479 ± 1.6103 mg/dL for the last, allowing a discrimination between both groups. Results were validated against a UV-Vis-NIR spectrophotometer with a correspondence of R² of 0.98194 between both instruments. Results suggest the potential application of the developed device to the sensitive detection of relevant analytes with a low-cost, user-friendly, low-power and portable instrumentation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5713636PMC
http://dx.doi.org/10.3390/s17112495DOI Listing

Publication Analysis

Top Keywords

evaluation low-cost
8
low-cost colorimeter
8
salivary samples
8
glucose
5
colorimeter glucose
4
detection
4
glucose detection
4
detection salivary
4
samples limited
4
limited access
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!