Influence of hydrostatic pressure on intracoronary indices of stenosis severity in vivo.

Clin Res Cardiol

Department of Cardiology, Klinikum Oldenburg, European Medical School Oldenburg-Groningen, Carl von Ossietzky University Oldenburg, Rahel-Straus-Str. 10, 26133, Oldenburg, Germany.

Published: March 2018

Background: An influence of hydrostatic pressure on intracoronary indices of stenosis severity in vitro was recently reported. We sought to analyze the influence of hydrostatic pressure, caused by the height difference between the distal and proximal pressure sensor after guidewire positioning in the interrogated vessel, on intracoronary pressure measurements in vivo.

Methods And Results: In 30 coronary stenoses, intracoronary pressure measurements were performed in supine, left, and right lateral patient position. Height differences between the distal and proximal pressure sensor were measured by blinded observers. Measurement results of the position with the highest ("high") and lowest height difference ("low") were compared. In group "high", all measured indices were higher: mean difference of fractional flow reserve (FFR) 0.045 (SD 0.033, 95% CI 0.033-0.057, p < 0.0001), of instantaneous wave-free ratio (iFR) 0.043 (SD 0.04, 95% CI 0.029-0.057, p < 0.0001), and of resting Pd/Pa 0.037 (SD 0.034, 95% CI 0.025-0.049, p < 0.0001). Addition of the physically expectable hydrostatic pressure to the distal coronary pressures of the control group abolished the differences: corrected ∆FFR - 0.006 (SD 0.027, 95% CI - 0.015 to 0.004, p = 0.26), corrected ∆Pd/Pa - 0.008 (SD 0.03, 95% CI - 0.019 to 0.003, p = 0.18). Adjustment for hydrostatic pressure of FFR values in a standard supine position increased all values in anterior vessels and decreased all values in posterior vessels. The mean changes of FFR due to adjustment were: LAD - 0.048 (SD 0.016), CX 0.02 (SD 0.009), RCA 0.02 (SD 0.021). Dichotomous severity classification changed in 12.9% of stenoses.

Conclusions: The study demonstrates a relevant influence of hydrostatic pressure on intracoronary indices of stenosis severity in vivo, caused by the height differences between distal and proximal pressure sensor.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00392-017-1174-2DOI Listing

Publication Analysis

Top Keywords

influence hydrostatic
12
hydrostatic pressure
12
pressure intracoronary
8
intracoronary indices
8
indices stenosis
8
stenosis severity
8
height difference
8
distal proximal
8
proximal pressure
8
pressure sensor
8

Similar Publications

Structural and physicochemical properties of debranched lotus seed starch treated with high hydrostatic pressure.

Int J Biol Macromol

December 2024

College of Food Science, Fujian Agriculture and Forestry University, Fuzhou, Fujian, PR China; Fujian Provincial Key Laboratory of Quality Science and Processing Technology in Special Starch, Fuzhou, Fujian, PR China. Electronic address:

Lotus seeds represent a significant economic crop and are abundant in starch. To further enhance their application value, this study investigates the structural characteristics of lotus seed starch (LS) under the combined influence of pullulanase and high hydrostatic pressure (HHP). Pullulanase increased amylose content from 39.

View Article and Find Full Text PDF

High hydrostatic pressure modulates the digestive properties of rice starch-gallic acid composites by boosting non-inclusion complexation.

Int J Biol Macromol

December 2024

Department of Applied Sciences, Northumbria University, Newcastle Upon Tyne NE1 8ST, Tyne & Wear, England, United Kingdom. Electronic address:

Influencing the starch postprandial glycemia via interventions that are sourced from natural plant materials has gained attention recently. Amylose present in starch is reported to form complexes with small ligands such as gallic acid (GA) through a conformational change that are digested slowly and contribute to the formation of resistant starch. In this study, the molecular interactions, multi-scale structure and in vitro digestion properties of normal neat rice starch and rice starch-GA composites (2, 5 % w/v) obtained either by high hydrostatic pressure (HHP) or thermal (T) treatment were compared.

View Article and Find Full Text PDF

Study on numerical simulation of groundwater flow field and slope stability in multi-aquifer open pit mine.

Sci Rep

December 2024

Liaoning Institute of Technology and Equipment for Mineral Resources Development and Utilisation in Higher Educational Institutions, Liaoning Technical University, Fuxin, 123000, Liaoning, China.

Water is one of the most important influences on slope stability in open pit mines. In order to solve the problem of slope stability analysis in multi-aquifer open pit mines, the open pit mine in Block I of Thar Coalfield in Pakistan with multiple aquifers was taken as the research background. The groundwater flow field at different excavation phases was analyzed by numerical simulation method.

View Article and Find Full Text PDF

The dynamic behavior of articular cartilage (a soft porous biological tissue) with strain-dependent nonlinear permeability under cyclic compressive loading and magnetic field is investigated computationally. The compressive force is applied on top surface of the cylindrical plug of the tissue by means of a porous filter. The study of mechanical and deformational behavior of soft porous tissues such as articular cartilage under dynamic compressive loading and magnetic field is useful in understanding the underlying mechano-biological process that may lead to the development of a treatment and recovery protocol in a diseased state.

View Article and Find Full Text PDF

The dynamic control of chiral (enantiomeric) responses in chiral host-guest complexes through external stimuli is a significant challenge in modern chemistry for developing smart stimuli-responsive materials. Herein, we report the (chir)optical properties and chiral recognition behavior of water-soluble chiral naphthotubes () under the influence of hydrostatic pressure as an external stimulus. The hydrostatic pressure spectral profiles compared to those obtained at normal pressure revealed the dynamic behavior of under hydrostatic pressure, owing to the flexible linker.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!