Model for Estimating Power and Downtime Effects on Teletherapy Units in Low-Resource Settings.

J Glob Oncol

, , , , , , , and , The University of Texas MD Anderson Cancer Center; The University of Texas Graduate School of Biomedical Sciences at Houston, Houston, TX; , Benavides Cancer Institute, University of Santo Tomas Hospital, Manila, Philippines; , University of Alabama Birmingham, Birmingham, AL; , American University of Beirut Medical Center, Beirut, Lebanon; and , Groote Schuur Hospital, Cape Town, South Africa.

Published: October 2017

Purpose: More than 6,500 megavoltage teletherapy units are needed worldwide, many in low-resource settings. Cobalt-60 units or linear accelerators (linacs) can fill this need. We have evaluated machine performance on the basis of patient throughput to provide insight into machine viability under various conditions in such a way that conclusions can be generalized to a vast array of clinical scenarios.

Materials And Methods: Data from patient treatment plans, peer-reviewed studies, and international organizations were combined to assess the relative patient throughput of linacs and cobalt-60 units that deliver radiotherapy with standard techniques under various power and maintenance support conditions. Data concerning the frequency and duration of power outages and downtime characteristics of the machines were used to model teletherapy operation in low-resource settings.

Results: Modeled average daily throughput was decreased for linacs because of lack of power infrastructure and for cobalt-60 units because of limited and decaying source strength. For conformal radiotherapy delivered with multileaf collimators, average daily patient throughput over 8 years of operation was equal for cobalt-60 units and linacs when an average of 1.83 hours of power outage occurred per 10-hour working day. Relative to conformal treatments delivered with multileaf collimators on the respective machines, the use of advanced techniques on linacs decreased throughput between 20% and 32% and, for cobalt machines, the need to manually place blocks reduced throughput up to 37%.

Conclusion: Our patient throughput data indicate that cobalt-60 units are generally best suited for implementation when machine operation might be 70% or less of total operable time because of power outages or mechanical repair. However, each implementation scenario is unique and requires consideration of all variables affecting implementation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5646876PMC
http://dx.doi.org/10.1200/JGO.2016.005306DOI Listing

Publication Analysis

Top Keywords

cobalt-60 units
20
patient throughput
16
teletherapy units
8
low-resource settings
8
power outages
8
average daily
8
delivered multileaf
8
multileaf collimators
8
units
7
throughput
7

Similar Publications

Errors in manual radiotherapy treatment procedures and their evolution in a low resource setting: Uganda's experience.

Phys Med

February 2024

Medical Technology, Health Information and Research Directorate, Ministry of Health P.O. Box 1176 Jerusalem, Israel.

Purpose: In Uganda, two-dimensional (2D) radiotherapy treatments have been in use since the establishment of radiotherapy in 1995. Preliminary investigations of treatment records in November 2019 showed evidence of gaps requiring urgent attention. The purpose of this study was to improve the safety of the treatments.

View Article and Find Full Text PDF

The inequitable access to radiotherapy globally is a complex undertaking that will require sustained work identifying gaps and mobilising efforts to resolve. The purpose of this review is to identify gaps and needs in radiotherapy in Latin America and the Caribbean. Data from 41 countries in Latin America and the Caribbean on teletherapy megavoltage units and brachytherapy resources were extracted from the International Atomic Energy Agency (IAEA) Directory of Radiotherapy Centers.

View Article and Find Full Text PDF

Background: In a recent study, we reported beam quality correction factors, f , in carbon ion beams using Monte Carlo (MC) methods for a cylindrical and a parallel-plate ionization chamber (IC). A non-negligible perturbation effect was observed; however, the magnitude of the perturbation correction due to the specific IC subcomponents was not included. Furthermore, the stopping power data presented in the International Commission on Radiation Units and Measurements (ICRU) report 73 were used, whereas the latest stopping power data have been reported in the ICRU report 90.

View Article and Find Full Text PDF

The threat of a nuclear attack has increased in recent years highlighting the benefit of developing additional therapies for the treatment of victims suffering from Acute Radiation Syndrome (ARS). In this work, we evaluated the impact of a PEGylated thrombopoietin mimetic peptide, JNJ-26366821, on the mortality and hematopoietic effects associated with ARS in mice exposed to lethal doses of total body irradiation (TBI). JNJ-26366821 was efficacious as a mitigator of mortality and thrombocytopenia associated with ARS in both CD2F1 and C57BL/6 mice exposed to TBI from a cobalt-60 gamma-ray source.

View Article and Find Full Text PDF

Introduction: Gamma Irradiation of blood products accentuates biochemical changes in the blood stored at 4°C. This study tried to compare the changes in potassium, sodium, glucose, lactate, and lactate dehydrogenase (LDH) levels in packed red blood cell (PRBC) units irradiated at various time points and then stored versus those stored for a particular period then irradiated.

Methodology: One hundred and eighty units of RBCs were randomly assigned equally to be irradiated or not.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!