A highly versatile automatized setup for quantitative measurements of PHIP enhancements.

J Magn Reson

Technische Universität Darmstadt, Eduard-Zintl-Institut für Anorganische und Physikalische Chemie, Alarich-Weiss-Straße 8, Darmstadt 64287, Germany. Electronic address:

Published: December 2017

The design and application of a versatile and inexpensive experimental extension to NMR spectrometers is described that allows to carry out highly reproducible PHIP experiments directly in the NMR sample tube, i.e. under PASADENA condition, followed by the detection of the NMR spectra of hyperpolarized products with high spectral resolution. Employing this high resolution it is feasible to study kinetic processes in the solution with high accuracy. As a practical example the dissolution of hydrogen gas in the liquid and the PHIP kinetics during the hydrogenation reaction of Fmoc-O-propargyl-l-tyrosine in acetone-d are monitored. The timing of the setup is fully controlled by the pulse-programmer of the NMR spectrometer. By flushing with an inert gas it is possible to efficiently quench the hydrogenation reaction in a controlled fashion and to detect the relaxation of hyperpolarization without a background reaction. The proposed design makes it possible to carry out PHIP experiments in an automatic mode and reliably determine the enhancement of polarized signals.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmr.2017.10.007DOI Listing

Publication Analysis

Top Keywords

phip experiments
8
hydrogenation reaction
8
highly versatile
4
versatile automatized
4
automatized setup
4
setup quantitative
4
quantitative measurements
4
phip
4
measurements phip
4
phip enhancements
4

Similar Publications

Hyphenation of 2D NMR With Hydrogenative PHIP.

Magn Reson Chem

January 2025

Institute for Molecules and Materials, Radboud University, Nijmegen, The Netherlands.

Parahydrogen induced polarisation (PHIP) is often used to enhance the sensitivity of NMR, with the purpose of extending the applicability of the technique. Nuclear spin hyperpolarisation obtained via PHIP is generally localised on the protons derived from the addition of para-enriched hydrogen to an unsaturated substrate. This limitation has been previously addressed by pulse schemes that can spread this hyperpolarised magnetisation through the entire network of J-coupled protons in the product molecule.

View Article and Find Full Text PDF

Background: Multiple sclerosis (MS) is a neuroinflammatory disease triggered by a combination of genetic traits and external factors. Autoimmune nature of MS is proven by the identification of pathogenic T cells, but the role of autoantibody-producing B cells is less clear. A comprehensive understanding of the development of neuroinflammation and the identification of targeted autoantigens are crucial for timely diagnosis and appropriate treatment.

View Article and Find Full Text PDF

: Ocular predominant mucous membrane pemphigoid (oMMP) is a severe subtype of autoimmune blistering disease (AIBD), which can result in scarring and vision loss. The diagnosis of oMMP is challenging as patients often have undetectable levels of circulating autoantibodies by conventional assays. Likewise, the principal autoantigen in oMMP has been an area of debate.

View Article and Find Full Text PDF

Radio Amplification by Stimulated Emission of Radiation (RASER) is a phenomenon observed during nuclear magnetic resonance (NMR) experiments with strongly negatively polarized systems. This phenomenon may be utilized for the production of very narrow NMR lines, background-free NMR spectroscopy, and excitation-free sensing of chemical transformations. Recently, novel methods of producing RASER by ParaHydrogen-Induced Polarization (PHIP) were introduced.

View Article and Find Full Text PDF

Alterations in pH are a hallmark in several pathologies including cancer, ischemia, and inflammation. Non-invasive magnetic resonance methods to measure pH offer a new approach for early diagnosis of diseases characterized by acid-base imbalances. The hyperpolarization with parahydrogen-induced polarization (PHIP) enhances inherently low signals in magnetic resonance experiments by several orders of magnitude and offers a suitable platform to obtain biocompatible markers in less than one minute.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!