Objectives: The aim of this study was to evaluate the hemodynamic performance of a sutureless bioprosthesis under high workload at mid-term follow-up.

Methods: Thirty-two patients who underwent isolated aortic valve replacement with a Perceval sutureless bioprosthesis with a minimum follow-up of 1 year were enrolled in this study. S size prosthesis was deployed in 10 patients (31.3%), M size in 9 (28.1%), L size in 8 (25%) and XL size in 5 (15.6%). Effective orifice area (EOA), EOA index (EOAi), and transvalvular gradients were assessed at rest and during dobutamine stress echocardiography (DSE) a median of 19.5 months after surgery.

Results: Dobutamine stress echocardiography (DSE) significantly increased heart rate, stroke volume, ejection fraction, and transvalvular gradients (peak gradient, 24.0 ± 7.6 vs 38.7 ± 13.6 mm Hg, P < .001; mean gradient, 12.6 ± 4.2 vs 19.8 ± 8.3, P < .001). When compared to baseline, estimated valve areas significantly increased at follow-up (EOA, 1.48 ± 0.46 vs 2.06 ± 0.67, P < .001; EOAi, 0.84 ± 0.26 vs 1.17 ± 0.37, P < .001). Mean percentage increase in EOAi was 40.3% ± 28.0%. S size prostheses had the highest increase in EOA1, but the difference was not significant (S 46.0% ± 27.5% vs M 45.4% ± 34.5% vs L 32.7% ± 26.4% vs XL 32.1% ± 20.5%, P = .66). Severe patient-prosthesis mismatch (EOAi ≤ 0.65 cm /m ) was present at rest in 8 patients (25%), but only in one patient (3.1%) during DSE.

Conclusions: The Perceval sutureless bioprosthesis demonstrated good hemodynamics at rest and under high workload. The significant increase in EOAi during DSE suggests the potential advantages of Perceval sutureless bioprostheses in case of small aortic annulus or when patient-prosthesis mismatch is anticipated.

Download full-text PDF

Source
http://dx.doi.org/10.1111/echo.13735DOI Listing

Publication Analysis

Top Keywords

sutureless bioprosthesis
12
dobutamine stress
12
stress echocardiography
12
perceval sutureless
8
transvalvular gradients
8
echocardiography dse
8
hemodynamic assessment
4
assessment perceval
4
bioprosthesis dobutamine
4
echocardiography objectives
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!