Capillarys 2 Flex Piercing Detected a Rare Case of Hb Broomhill.

Ann Clin Lab Sci

Department of Laboratory medicine, West China Hospital, Sichuan University, Chengdu, P.R. China

Published: September 2017

We report a case of an extremely rare hemoglobin (Hb) variant-Hb Broomhill, which has been only reported once in the literature. Hemoglobin fractions were determined by capillary electrophoresis (Sebia Capillarys 2 Flex piercing) and high performance liquid chromatography (HPLC) (Bio-Rad Variant™ II Hemoglobin Testing System), respectively. Complete blood count and DNA sequencing were also performed. The capillary electrophoregram revealed a tiny shoulder peak before the HbA peak and a subtle abnormal HbA peak (slightly wider and lower), even though the percentage of each hemoglobin fraction was within the reference range (HbA, 97.4%; HbA, 2.6%). On HPLC, not only the percentage but also the peak shape of each hemoglobin fraction was normal (HbA 88.2%, HbA 2.5%, HbF 0.6%). Eventually, sequencing analysis of α genes confirmed a missense mutation (CCC>GCC at codon 114 in alpha1 gene) which caused Hb Broomhill variant. Our report suggest that capillary electrophoresis may be an accurate tool for screening and diagnosis of Hb disorders.

Download full-text PDF

Source

Publication Analysis

Top Keywords

capillarys flex
8
flex piercing
8
capillary electrophoresis
8
hba peak
8
hemoglobin fraction
8
hba
6
hemoglobin
5
piercing detected
4
detected rare
4
rare case
4

Similar Publications

A significant limitation of imaged capillary electric focusing (icIEF) is the inability to identify and characterize specific species in the electropherogram. This has led to the development of complementary ion-exchange chromatography (IEX)-based methods that are amenable to either fraction collection and subsequent characterization or online IEX coupled to mass spectrometry. To overcome this limitation while maintaining the use of icIEF, novel approaches, including an icIEF separation and fractionation technology (MauriceFlex, ProteinSimple), have been developed.

View Article and Find Full Text PDF

Objectives: In the field of Laboratory Medicine, the evolution of knowledge and the innovation of technologies are the basis of analytical and diagnostic progress, leading to the development of new solutions based on innovative technologies. However, these advances must be accompanied by evidence of appropriateness, diagnostic effectiveness, and organizational efficiency, considering the test's first impact on patient outcomes.

Methods: The Health Technology Assessment (HTA) is a valid management tool to support Laboratory Medicine professionals in assessing technologies and which is the most appropriate to adopt.

View Article and Find Full Text PDF

Background: Haemoglobin (Hb) variants such as sickle cell trait (SCT/HbAS) play a role in protecting against clinical malaria, but little is known about the development of immune responses against malaria parasite (Plasmodium falciparum surface protein 230 (Pfs230) and Plasmodium falciparum erythrocyte binding antigen 175 region-3 (PfEBA175-3R)) and vector (on the An. gambiae Salivary Gland Protein-6 peptide 1 (gSG6-P1)) antigens in individuals with variants Hb genotypes. This study assessed antibody (IgG) responses against malaria parasite, Pfs230 and PfEBA175-3R and vector, gSG6-P1 in febrile individuals with variant Hb genotypes.

View Article and Find Full Text PDF

Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test.

Pract Lab Med

January 2024

Emergency Medicine Research Group (EMERGE), Royal Infirmary of Edinburgh, Edinburgh, UK.

C-reactive protein (CRP) is an established acute-phase marker for infection, inflammation and tissue injury, used to guide clinical decision-making in primary and secondary care. This study compared the analytical performance of the quantitative microfluidic point-of-care LumiraDx CRP Test to a laboratory-based reference method (Siemens RCRP Flex assay on the Dimension® Xpand®) and evaluated equivalence of sample matrices (blood versus plasma) in point-of-care settings using samples from patients presenting with symptoms of infection or inflammation. The LumiraDx CRP Test demonstrated close agreement with the lab reference test (range, 5.

View Article and Find Full Text PDF

Traditional Western blotting is one of the most used analytical techniques in biological research. However, it can be time-consuming and suffer from a lack of reproducibility. Consequently, devices with different degrees of automation have been developed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!