Comparison of Active Stimulating Electrodes of Sacral Neuromodulation.

Neuromodulation

Medtronic, Inc., Restorative Therapies Group, Research & Core Technology, Minneapolis, MN, USA.

Published: December 2017

Objective: The goal of this study was to compare the motor response to sacral neuromodulation (SNM) with different pairs of stimulating electrodes in anesthetized and awake sheep.

Materials And Methods: Similar to SNM clinical use in humans, the InterStim quadripolar tined lead was implanted adjacent to the S3 nerve root in sheep and bipolar stimulation was configured with one electrode negative and one electrode positive on the four contacts (0 most distal to device, 1, 2, and 3 most proximal).

Results: Electrode 3-cathode and electrode 0-anode (3-/0+) stimulation had the lowest visual response threshold (0.46 ± 0.14V, anesthetized, 0.56 ± 0.21V, conscious), representing the most sensitive stimulation. Stimulation on electrode 0 (0-/1+) had the highest response threshold among tested electrodes (2.70 ± 0.23V, anesthetized, 3.38 ± 0.96V, conscious). The order according to response threshold from low to high was 3 < 2 < 1 < 0. The triggered response by 3-/0+ stimulation solely occurred in the perineum, tail, or bellows. In contrast, the 0-/1+ stimulation frequently evoked response in gluteal and thigh regions. The electromyographic activities from the anus were sensitive to low intensities of stimulation on electrode 3 (e.g., 3-/0+, 3-/2+).

Conclusions: Objective motor responses to SNM as a functional indicator for optimal lead placement may be used to demonstrate that the contact which is most proximal to the foramen (electrode 3) is an optimal electrode to trigger an "on-target" response to lower intensity stimulation. Data from this preclinical work suggest that there are several principles that may be referenced to simplify and expedite the programming process in clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1111/ner.12710DOI Listing

Publication Analysis

Top Keywords

response threshold
12
stimulating electrodes
8
sacral neuromodulation
8
electrode
5
comparison active
4
active stimulating
4
electrodes sacral
4
neuromodulation objective
4
objective goal
4
goal study
4

Similar Publications

Glucose metabolism and hormonal responses are largely affected by exercise intensity, which exhibits an alternating pattern in many sports activities. The aim of this study was to investigate and compare glycemic and hormonal responses during and after exercise of constant (CON) and alternating (ALT) intensity with the same duration and total work. Ten healthy male volunteers performed two, 60 min cycling exercise bouts in random order: the ALT bout, where exercise intensity alternated between 46.

View Article and Find Full Text PDF

Plastic Creep Constraint in Nylon Instrument Strings.

Materials (Basel)

January 2025

Department of Engineering, University of Cambridge, Trumpington St, Cambridge CB2 1PZ, UK.

A number of rectified nylon harp strings, having the same nominal diameter, were subjected to different sequences of applied stress steps. Each string was tested continuously for several weeks to allow sufficient time for the stretching responses to be clearly observed. Qualitatively, much of the observed behaviour was in accordance with established expectations.

View Article and Find Full Text PDF

Functional and Structural Changes in the Inner Ear and Cochlear Hair Cell Loss Induced by Hypergravity.

Int J Mol Sci

January 2025

Department of Otorhinolaryngology-Head and Neck Surgery, College of Medicine, Inha University, Incheon 22332, Republic of Korea.

Gravitational changes have been shown to cause significant abnormalities in various body systems, including the cardiovascular, immune, vestibular, and musculoskeletal systems. While numerous studies have examined the response of the vestibular system to gravitational stimulation, research on functional changes in the peripheral inner ear remains limited. The inner ear comprises two closely related structures: the vestibule and cochlea.

View Article and Find Full Text PDF

We report a low-cost, portable biosensor composed of an aptamer-functionalized nanoporous anodic aluminum oxide (NAAO) membrane and a commercial microcontroller chip-based impedance reader suitable for electrochemical impedance spectroscopy (EIS)-based sensing. The biosensor consists of two chambers separated by an aptamer-functionalized NAAO membrane, and the impedance reader is utilized to monitor transmembrane impedance changes. The biosensor is utilized to detect amodiaquine molecules using an amodiaquine-binding aptamer (OR7)-functionalized membrane.

View Article and Find Full Text PDF

This study investigates the effects of negative bias temperature (NBT) stress and irradiation on the threshold voltage () of p-channel VDMOS transistors, focusing on degradation, recovery after each type of stress, and operational behavior under varying conditions. Shifts in (Δ) were analyzed under different stress orders, showing distinct influence mechanisms, including defects creation and their removal and electrochemical reactions. Recovery data after each type of stress indicated ongoing electrochemical processes, influencing subsequent stress responses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!