Objectives: To evaluate the image quality and lesion conspicuity of virtual-monochromatic-imaging (VMI) with dual-layer DECT (DL-DECT) for reduced-iodine-load multiphasic-hepatic CT.
Methods: Forty-five adults with renal dysfunction who had undergone hepatic DL-DECT with 300-mgI/kg were included. VMI (40-70-keV, DL-DECT-VMI) was generated at each enhancement phase. As controls, 45 matched patients undergoing standard 120-kVp protocol (120-kVp, 600-mgI/kg, and iterative reconstruction) were included. We compared the size-specific dose estimate (SSDE), image noise, CT attenuation, and contrast-to-noise ratio (CNR) between protocols. Two radiologists scored the image quality and lesion conspicuity.
Results: SSDE was significantly lower in DL-DECT group (p < 0.01). Image noise of DL-DECT-VMI was almost constant at each keV (differences of ≤15%) and equivalent to or lower than of 120-kVp. As the energy decreased, CT attenuation and CNR gradually increased; the values of 55-60 keV images were almost equivalent to those of standard 120-kVp. The highest scores for overall quality and lesion conspicuity were assigned at 40-keV followed by 45 to 55-keV, all of which were similar to or better than of 120-kVp.
Conclusions: For multiphasic-hepatic CT with 50% iodine-load, DL-DECT-VMI at 40- to 55-keV provides equivalent or better image quality and lesion conspicuity without increasing radiation dose compared with standard 120-kVp protocol.
Key Points: • 40-55-keV yields optimal image quality for half-iodine-load multiphasic-hepatic CT with DL-DECT. • DL-DECT protocol decreases radiation exposure compared with 120-kVp scans with iterative reconstruction. • 40-keV images maximise conspicuity of hepatocellular carcinoma especially at hepatic-arterial phase.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00330-017-5114-3 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!