Dual high and low energy images of Dual Energy X-ray Absorptiometry (DEXA) suffer from noises due to the use of weak amount of X-rays. Denoising these DEXA images could be a key process to enhance and improve a Bone Mineral Density (BMD) map which is derived from a pair of high and low energy images. This could further improve the accuracy of diagnosis of bone fractures, osteoporosis, and etc. In this paper, we present a denoising technique for dual high and low energy images of DEXA via non-local means filter (NLMF). The noise of dual DEXA images is modeled based on both source and detector noises of a DEXA system. Then, the parameters of the proposed NLMF are optimized for denoising utilizing the experimental data from uniform phantoms. The optimized NLMF is tested and verified with the DEXA images of the uniform phantoms and real human spine. The quantitative evaluation shows the improvement of Signal-to-Noise Ratio (SNR) for the high and low phantom images on the order of 30.36% and 27.02% and for the high and low real spine images on the order of 22.28% and 33.43%, respectively. Our work suggests that denoising via NLMF could be a key preprocessing process for clinical DEXA imaging.

Download full-text PDF

Source
http://dx.doi.org/10.1109/EMBC.2017.8036889DOI Listing

Publication Analysis

Top Keywords

high low
20
dexa images
16
low energy
12
energy images
12
images
9
non-local filter
8
dexa
8
denoising dexa
8
images dual
8
dual high
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!