2D-LC-MS/MS to measure cleaved high-molecular-weight kininogen in human plasma as a biomarker for C1-INH-HAE.

Bioanalysis

Bioanalytical & Biomarker Development, Research & Nonclinical Development, Shire, Lexington, MA 02421, USA.

Published: October 2017

Aim: C1-INH-HAE is caused by activation of plasma kallikrein which subsequently cleaves high-molecular-weight kininogen (HMWK) to generate bradykinin and cHMWK.

Materials & Methods: A novel ion-pair 2D LC-MS/MS assay was developed to measure the 46 kDa cHMWK in plasma as a biomarker for C1-INH-HAE. The sample preparation included sodium dodecyl sulfate denaturation, methanol crash, chymotryptic digestion and peptide enrichment by solid phase extraction.

Results: The LLOQ was 200 ng/ml. The overall cHMWK recovery combining crash and digestion was 57.5%. The precision of the method was ≤12.7% and accuracy ≤-13.8%.

Conclusion: A reagent-free LC-MS assay has been developed for the quantitation of 46 kDa cHMWK, which was shown to be elevated in plasma of C1-INH-HAE patients due to C1-INH deficiency relative to that of healthy subjects.