Accuracy of Oxygen Flow Delivered by Compressed-Gas Cylinders in Hospital and Prehospital Emergency Care.

Respir Care

Institut de Recherche Expérimentale et Clinique (IREC), Pôle de Pneumologie, ORL & Dermatologie, Service de Pneumologie, Université Catholique de Louvain, Brussels, Belgium.

Published: March 2018

Background: Oxygen cylinders are widely used both in hospital and prehospital care. Excessive or inappropriate F may be critical for patients with hypercapnia or hypoxia. Moreover, over-oxygenation could be deleterious in ischemic disorders. Supplemental oxygen from oxygen cylinder should therefore be delivered accurately. The aim of this study was to assess the accuracy of oxygen flows for oxygen cylinder in hospital and prehospital care.

Methods: A prospective trial was conducted to evaluate accuracy of delivered oxygen flows (2, 4, 6, 9 and 12 L/min) for different oxygen cylinder ready for use in different hospital departments. Delivered flows were analyzed randomly using a calibrated thermal mass flow meter. Two types of oxygen cylinder were evaluated: 78 oxygen cylinder with a single-stage regulator and 70 oxygen cylinder with a dual-stage regulator. Delivered flows were compared to the required oxygen flow. The residual pressure value for each oxygen cylinder was considered. A coefficient of variation was calculated to compare the variability of the delivered flow between the two types of oxygen cylinder.

Results: The median values of delivered flows were all ≥ 100% of the required flow for single stage (range 100-109%) and < 100% of required flow for dual stage (range 95-97%). The median values of the delivered flow differed between single and dual stage. It was found that single stage is significantly higher than dual stage ( = .01). At low flow, the dispersion of the measures for single stage was higher than with a high oxygen flow. Delivered flow differences were also found between low and high residual pressures, but only with single stage ( = .02). The residual pressure for both oxygen cylinders (no. = 148) ranged from 73 to 2,900 pounds per square inch, and no significant difference was observed between the 2 types ( = .86). The calculated coefficient of variation ranged from 7% (±1%) for dual stage to 8% (±2%) for single stage.

Conclusions: This study shows good accuracy of oxygen flow delivered via oxygen cylinders. This accuracy was higher with dual stage. Single stage was also accurate, however, at low flow this accuracy is slightly less. Moreover, with single stage, when residual pressure decreases, the median value of delivered flow decreased.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.05657DOI Listing

Publication Analysis

Top Keywords

oxygen cylinder
28
single stage
24
dual stage
20
oxygen
17
oxygen flow
16
delivered flow
16
flow
13
accuracy oxygen
12
flow delivered
12
hospital prehospital
12

Similar Publications

Respiratory interventions including noninvasive ventilation, continuous positive airway pressure and high-flow nasal oxygen generated infectious aerosols may increase risk of airborne disease (SARS-CoV-2, influenza virus) transmission to healthcare workers. We developed and tested a prototype portable UV-C device to sterilize high flows of viral-contaminated air from a simulated patient source at airflow rates of up to 100 l/m. Our device consisted of a central quartz tube surrounded 6 high-output UV-C lamps, within a larger cylinder allowing recirculation past the UV-C lamps a second time before exiting the device.

View Article and Find Full Text PDF

Alcohol fuels with different carbon numbers such as propanol (C3), butanol (C4) and pentanol (C5) have lately become popular in both conventionally and RCCI-operated diesel engines thanks to their high cetane number (CN) and oxygen content along with lower latent heat of evaporation, which are useful for reducing the high CO/HC emissions, whereas RCCI mode still suffers from these emissions. Therefore, in this study, these three alcohol fuels with carbon numbers ranging from C3 to C5 were employed as low-reactivity fuel (LRF) in a single-cylinder RCCI engine under a constant engine speed of 2400 rpm and varying loadings (from 20 to 60% of full load at 20% intervals) and premixed ratios (from 0 to 60% with 15% intervals) when using B7 as high-reactivity fuel (HRF). In the experimental study, the effect of oxygen content, latent heat of evaporation, and cetane number which changes linearly with the carbon number of alcohols used, on exhaust emissions, were analyzed.

View Article and Find Full Text PDF

Lankford, Harvey V, Hannah Kimberley. Supplemental Oxygen on the Annie Smith Peck Expedition of 1903 to 6,367 m (20,892 ft) Mount Sorata (Illampu). 00:00-00, 2024.

View Article and Find Full Text PDF

Cold atmospheric plasma (CAP) enables painless tissue treatment by producing reactive species including excited molecules and charged particles and is of great interest for medical applications. Medical CAP sources work in contact with air at ambient pressure, resulting in the generation of substantial amounts of reactive oxygen and nitrogen radicals. These radicals have a significant influence on cellular biochemistry, are crucial components of the immune system, and play a central role in wound therapy.

View Article and Find Full Text PDF

This study presents a gas chromatographic detector using alternating current (AC) discharged in air to generate μ-arc at atmosphereic pressure. This air-based μ-arc emission detector (μ-AED) was assembled by two stainless-steel syringe needles inside a quartz tube. The length of μ-arc (i.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!